Instrument Function
R&S
®
SMF100A
234
Operating Manual 1167.2319.02 ─ 12
"Duty Cycle"
Indicates the ratio between the pulse duration (
τ
) and the pulse period
(T) of the measured pulse signal in per cent:
Duty Cycle = (pulse duration / pulse period) * 100
Remote command:
:TRACe<ch>[:POWer]:SWEep:MEASurement:PULSe:DCYCle:DISPlay:
ANNotation[:STATe]
:TRACe<ch>[:POWer]:SWEep:MEASurement:PULSe:DCYCle?
"Pulse Width"
Indicates the pulse duration of the pulse data measurement in sec-
onds.
Remote command:
:TRACe<ch>[:POWer]:SWEep:MEASurement:PULSe:DURation:DISPlay:
ANNotation[:STATe]
:TRACe<ch>[:POWer]:SWEep:MEASurement:PULSe:DURation?
"Pulse Period"
Indicates the time the pulse signal needs to complete one cycle.
Remote command:
:TRACe<ch>[:POWer]:SWEep:MEASurement:PULSe:PERiod:DISPlay:
ANNotation[:STATe]
:TRACe<ch>[:POWer]:SWEep:MEASurement:PULSe:PERiod?
"Pulse Off
Time"
Determines the time the pulse signal is low, that means as long as
the signal level is below the proximal value.
Remote command:
:TRACe<ch>[:POWer]:SWEep:MEASurement:PULSe:SEParation:DISPlay:
ANNotation[:STATe]
:TRACe<ch>[:POWer]:SWEep:MEASurement:PULSe:SEParation?
Transition Times - Pulse Data Analysis
Selects the transition parameters of the pulse signal to be indicated below the mea-
surement diagram.
The R&S NRP-Z sensor searches for the first rising edge and the first falling edge in
the trace.
Note:
The "Rearrange" key or the key combination <STR+A> toggle between diagram
views.
RF Level