
Putting the System into Operation
R&S
®
ZN-Z3x
14
User Manual 1177.6381.02 ─ 03
Table 2-3: Inline Calibration Unit Errors
Error code
Error description
001
CAN bus error
002
Memory read/write error
003
Temperature error
004
24
V error
Do not connect any further Inline Calibration Unit R&S
ZN-Z32/Z33,
otherwise the calibration precision (effective system data) can no lon-
ger be guaranteed
005
8
V error
Calibration functionality degraded
006
5
V error
CAN bus functionality degraded, risk of system error
Start Cal
Button with integrated LED, allowing to start a calibration sweep.
The button is armed using the SCPI command
. Its current state
can be read using
.
2.3
Preparing the System
This section covers the basic steps to set up the Inline Calibration System
R&S
ZN-Z3x and to put it into operation.
Preparing the System
Содержание R&S ZN-Z3x
Страница 1: ...R S ZN Z3x Inline Calibration System User Manual User Manual 1177 6381 02 03 ÛÍß2 ...
Страница 4: ...Contents R S ZN Z3x 4 User Manual 1177 6381 02 03 ...
Страница 22: ...Putting the System into Operation R S ZN Z3x 22 User Manual 1177 6381 02 03 Connecting External Devices ...
Страница 50: ...Maintenance R S ZN Z3x 50 User Manual 1177 6381 02 03 Packing and storage ...