Putting the System into Operation
R&S
®
ZN-Z3x
11
User Manual 1177.6381.02 ─ 03
●
Chapter 2.3.5, "Switching the ICC Off"
CAN 1,2
Two independent CAN bus interfaces for control connection and
power supply of Inline Calibration Units, see
The CAN bus voltage supplied by the R&S
ZN-Z30 is +24
V with a
maximum current of 1.8
A. The Inline Calibration Units will report an error if the
supply voltage is too low (e.g. due to excessive bus length).
Table 2-1: CAN 1,2 pin assignment
Contact
Description
Value Range
1 (northwest)
CAN_HIGH
0 ... +5.5
V
2 (southwest)
CAN_LOW
0 ... +5.5
V
3 (southeast)
+24
V supply voltage
0 ... +24.5
V
4 (northeast)
Ground
0
V
The color of the LED above the connectors indicates the state of the respective
CAN bus:
●
red: bus not working (e.g. no Inline Calibration Unit connected) or in error
state
●
blue: bootloader mode (temporary state)
●
pink: calibration data alignment between Inline Calibration Units and Inline
Calibration Controller (temporary state)
●
green: error-free operation
2.1.2
Back Panel
The back panel of the Inline Calibration Controller is shown on the photo below.
Brief explanations on the controls and connectors can be found in
.
Inline Calibration Controller Tour
Содержание R&S ZN-Z3x
Страница 1: ...R S ZN Z3x Inline Calibration System User Manual User Manual 1177 6381 02 03 ÛÍß2 ...
Страница 4: ...Contents R S ZN Z3x 4 User Manual 1177 6381 02 03 ...
Страница 22: ...Putting the System into Operation R S ZN Z3x 22 User Manual 1177 6381 02 03 Connecting External Devices ...
Страница 50: ...Maintenance R S ZN Z3x 50 User Manual 1177 6381 02 03 Packing and storage ...