Information on the Current Version and History
Release Notes for Firmware 1171.5360.52
13
● No valid DME pulse will be generated, if DME NRP trigger normalization failed due
to missing connection between RF output and R&S® NRP-Z81
● Wrong FM deviation, if the DME modulation was active before
● DME modulation: Mode “Reply”, Trigger Mode “Ext. Power Sensor”: Trigger does
not work after changing the frequency
● DME Modulation: After execution of “Normalize Setup” in DME Analysis the system
will hang for several minutes.
1.13 Version 2.15.185.15
Released: October 2010
Fixed Issues
● Adjustment after firmware update is not working due to restrictions in the file
system. This effect could only show up when updating older instruments with rsu-
files. Newer instuments (shipped from mid 2009) are not affected.
● Entering new license keys may lead to misleading error messages.
● Remote commands *IDN? and *OPT? are not working correctly when code
emulation is active, i.e. SCPI language is not native ("SCPI").
1.14 Version 2.15.185.13
Released: October 2010
New Functionality
● Support of new wide screen display (this version is necessary for all SMA with
serial number higher than 103000)
● Remote control emulation Agilent E4428 and E4438 (ESG)
● Remote control emulation Agilent E8257 and E8267 (PSG)
● Remote control emulation Agilent E8663
● Remote control emulation Agilent N5161, N5162, N5181, N5182, N5183 (MXG)
● Remote control emulation Racal 9087
● Remote control emulation Aeroflex 2051/2052
● Option R&S® SMA-K28 (Power Analysis): New trigger modes “Internal” and
“External”
● Option R&S® SMA-K28 (Power Analysis): User specific level offset
● Option R&S® SMA-K28 (Power Analysis): Added mathematical functions and
reference trace