Setting Up the Instrument
R&S
®
RTO2000
99
Getting Started 1332.9719.02 ─ 08
6
Setting Up the Instrument
Basic setup procedures for the instrument are the following:
....................................................................................99
........................................................................................99
............................................................................... 100
..................................................................................... 100
6.1
Performing a Self-alignment
The self-alignment aligns the data from several input channels vertically and horizon-
tally to synchronize the timebases, amplitudes and positions. The self-alignment proc-
ess includes a basic hardware check.
Recommendation on performing the self-alignment:
●
When putting the instrument into operation for the first time
●
After a firmware update
●
Once a week
●
When major temperature changes occur (>
5°)
Warm-up and prepare the instrument
Make sure that the instrument has been running and warming up before you start the
self-alignment. The minimum warm-up time is indicated in the data sheet.
Remove the probes from the input connectors.
1. On the "File" menu, select "Selfalignment".
2. On the "Control" tab, tap "Start Alignment".
The alignment is performed, the process might take several minutes. A message
box informs you about the running process, wait until this message box closes.
The overall pass/fail result is shown in the "Overall alignment state" field. The
results of the individual alignment steps for each input channel are indicated in the
"Results" tab. This information is required if problems arise.
6.2
Aligning the Touchscreen
When the device is delivered, the touchscreen is initially calibrated. However, to ensure
that the touchscreen responds to the finger contact correctly, a touchscreen alignment
is required. Only users with administrator rights can fulfill this task.
Aligning the Touchscreen
Содержание R&S RTO2000 Series
Страница 1: ...R S RTO2000 Oscilloscope Getting Started Getting Started Version 08 1332971902 P C2...
Страница 6: ...Contents R S RTO2000 6 Getting Started 1332 9719 02 08...
Страница 34: ...Instrument Tour R S RTO2000 34 Getting Started 1332 9719 02 08 Keys and Controls...
Страница 72: ...Trying Out the Instrument R S RTO2000 72 Getting Started 1332 9719 02 08 Saving Data...
Страница 102: ...Setting Up the Instrument R S RTO2000 102 Getting Started 1332 9719 02 08 Adjusting Passive Probes...