Rohde & Schwarz R&S OSP-B157W8 Скачать руководство пользователя страница 39

Tour of the Extension Module

R&S

®

OSP

B157W8

(PLUS)

 R&S

®

OSP

B157WX

39

Getting Started 1178.5808.02 ─ 03

automatically adapted to the supply voltage. To turn the power on or off, put
the AC power switch into position I (on) or 0 (off).

Table 5-1: External connectors overview of the R&S

 

OSP-B157WX

Panel Label

Direction Signal

Type

Value range

Front

DUT1 ..

. DUT8

BiDir

Device under test, channel

1 to 8, max. 30

 

dBm

RF K-Type

1 GHz to 40 GHz

Front

OUT1 ..

. OUT8

BiDir

Forwarding DUT1 to 8 to

the base module

RF SMA

1 GHz to 6 GHz

Front

X1

BiDir

Companion device (from

the base module)

RF SMA

1 GHz to 6 GHz

Front

Rx

Out

Spectrum analyzer / test

receiver

RF SMA

1 GHz to 40 GHz

Front

RxOUT In

Spectrum analyzer / test

receiver (from the base

module)

RF SMA

1 GHz to 6 GHz

Front

USB (2x, only R&S

OSP120)

disabled

 

 

Front

MONITOR DIGITAL

(only R&S OSP120)

disabled

 

 

Rear

LAN

BiDir

Ethernet

Digital RJ45

 

Rear

CAN

 

disabled

 

 

Note that 

"RxOUT"

 is - in spite of its name - an 

input port

 for signals com-

ing from the "Rx" port of the R&S

 

OSP-B157W8 or R&S

 

OSP-

B157W8PLUS base module. The label "RxOUT" was chosen due to the
connection with the Rx output port of the base module.

Connecting Devices and a Base Module to the Extension Module

Содержание R&S OSP-B157W8

Страница 1: ...R S OSP B157W8 PLUS R S OSP B157WX 8 Port Modules for Switch Units Getting Started Getting Started Version 03 1178580802 82...

Страница 2: ...later 2019 Rohde Schwarz GmbH Co KG M hldorfstr 15 81671 M nchen Germany Phone 49 89 41 29 0 Fax 49 89 41 29 12 164 Email info rohde schwarz com Internet www rohde schwarz com Subject to change Data...

Страница 3: ...uctions 10 2 6 R S OSP on the Internet 10 3 Preparing for Use 13 3 1 Unpacking and Checking 13 3 2 Mounting the Base Module 13 3 3 Mounting the Extension Module 19 3 4 Installing or Updating the Firmw...

Страница 4: ...Contents R S OSP B157W8 PLUS R S OSP B157WX 4 Getting Started 1178 5808 02 03...

Страница 5: ...ccording to the standards FCC 15 407 ETSI EN 300 328 ETSI EN 301 893 ETSI EN 302 502 ETSI EN 300 440 and others Intended use The R S OSP B157W8 R S OSP B157W8PLUS and R S OSP B157WX are designated for...

Страница 6: ...ed for in band use only The lower end of the frequency range of the R S OSP B157W8 and R S OSP B157W8PLUS depends on the selected paths and application It starts for example approximately at 100 MHz i...

Страница 7: ...for an extended attenuation range see Fig ure 4 3 Each of the 8 configurable channels in the R S OSP B157W8 and R S OSP B157W8PLUS has the following integrated semiconductor components An individually...

Страница 8: ...X forwards the DUT s signals it also switches the sig nal analyzer path from its RxOUT port input port to its Rx port This Rx port is connected to the signal analyzer or test receiver while the Rx por...

Страница 9: ...all functions of the modules It includes the contents of the getting started manual and provides explanations step by step procedures figures and examples The user manual supports you during your firs...

Страница 10: ...test cases that you can use in tests according to various FCC and ETSI standards Examples are FCC 15 407 ETSI EN 300 328 or ETSI EN 301 893 The software which requires a license is available at www ro...

Страница 11: ...ptions and their order numbers and optional accessories The brochure provides an overview of the instrument and deals with the specific characteristics See www rohde schwarz com brochure datasheet osp...

Страница 12: ...z com soft ware osp 2 6 5 Service Manual Describes handling failed modules module replacement troubleshooting and special remote control commands for service purposes The document also con tains spare...

Страница 13: ...r you can use the material to protect the control elements and con nectors Mounting modules into switch units If you have received one or more modules that are already integrated in switch units you c...

Страница 14: ...safety instructions for units with a removable casing Carefully follow the step by step instructions for module replacement to avoid injury and to help ensure safe operation Risk of injuries When remo...

Страница 15: ...pt for a module in the rear slot A11 5 If the rear slot A11 of the switch unit holds a module we recommend removing it Optionally you can leave it there but you cannot operate it due to missing contro...

Страница 16: ...ove the rear panel strut between slot A12 and slot A13 21 Screw the rear plate of the module to the switch unit s rear panel 10 counter sunk screws type Torx M2 5x8 included in the delivery step 34 us...

Страница 17: ...the opened switch unit before inserting it 27 Do not yet insert the module into the switch unit 28 Insert the single end of cable W20 with the 12 pole connector through the open front panel of the sw...

Страница 18: ...to connector X12 of the switch unit s con trol board 9 4 pole connector at the short end of cable W20 4 wires plugged into connector X201 on the module s RF board see detail at left hand side 10 4 pol...

Страница 19: ...wo slots rear slots A12 and A13 together with front slots A12F and A13F Typically the rear slot A11 remains free but optionally it can remain occupied by a single slot module not operational Only one...

Страница 20: ...eet back on immediately after mounting the casing Do not move the switch unit with the rear feet missing Removing an existing old R S OSP B157 module If you want to install the R S OSP B157WX in a swi...

Страница 21: ...blind plates cover unused front slots A12F or A13F of the switch unit remove these blind plates 11 Place the switch unit on its front handles 12 Optionally if at the switch unit s feet there are yell...

Страница 22: ...ix the module to the adapter plate 3 Top view of the complete module with the same 11 screws highlighted as in the isolated presentation 2 21 Insert the module into the switch unit from the front side...

Страница 23: ...WX and the main board of the switch unit Interface board connector Cable Main board connector X15 power input 6 wire cable X12 power supply X13 interface slot A12 26 pin ribbon cable X4 interface slot...

Страница 24: ...e unit 34 Reconnect the power cable to the unit 3 4 Installing or Updating the Firmware For installing or updating the firmware of the R S OSP B157W8 or R S OSP B157W8PLUS use the R S WMS32 software p...

Страница 25: ...wer LED flashes green When the installation or update is completed the base module restarts and is ready for use 3 5 Putting into Operation For using the base module and the extension module in the R...

Страница 26: ...Observe the following instructions when stacking instruments Never stack more than three instruments If you need to stack more than three instruments install them in a rack The overall load imposed o...

Страница 27: ...verloaded The overall load on the fol ded out feet must not exceed 100 N Fmax Risk of instrument damage due to insufficient airflow in a rack If you mount several instruments in a rack you need an eff...

Страница 28: ...witch unit cannot recognize the module Instead the module s LAN interface must be connected to the control computer and the switching configuration must be saved as a Hardware Setup in the R S WMS32 s...

Страница 29: ...2 92 mm K type RF connector adapters as port savers on the original connectors A set of 8 adapters is included in the delivery Fix the adapters with a torque of 0 8 Nm to 1 1 Nm Leave the port saver a...

Страница 30: ...ust be installed on your control computer Operation is described in the module s user manual and in the operating manual of the R S OSP switch units see chapter on R S EMC32 software On the front pane...

Страница 31: ...diagram Figure 4 3 Circuit diagram of the R S OSP B157W8 and R S OSP B157W8PLUS RF board Orange right Front DUT and instrument ports Orange left Instrument or trigger ports and data acquisition board...

Страница 32: ...and depends on the appli cable test standard This flexibility allows setting up and handling complex test and measurement sequences with the connected devices 4 1 Connecting Devices to the Base Module...

Страница 33: ...he supply voltage To turn the power on or off put the AC power switch into position I on or 0 off The following table describes all other external connectors of the base module Table 4 1 External conn...

Страница 34: ...3 V low 0 V If you use the base module R S OSP B157W8 or R S OSP B157W8PLUS in combination with the extension module R S OSP B157WX make the following changes to the connections listed above Connect...

Страница 35: ...e operation by the software R S OSP Panel Operation is described in the module s user manual and in the operating manual of the R S OSP switch units On the front panel the R S OSP B157WX has the conne...

Страница 36: ...rs see Chapter 5 1 The R S OSP B157WX operates according to the following circuit diagram Figure 5 3 Circuit diagram of the R S OSP B157WX RF board Orange right 8 DUT input and output ports front conn...

Страница 37: ...9 rack Observe maximum input power and no DC voltage The R S OSP B157WX typically forwards RF signals from a DUT to the R S OSP B157W8 or R S OSP B157W8PLUS Hence if you connect the R S OSP B157WX to...

Страница 38: ...es upper R S OSP B157WX lower R S OSP B157W8 or R S OSP B157W8PLUS 7 Connect the port Rx of the R S OSP B157WX to the input port of your mea surement instrument for example a spectrum analyzer or test...

Страница 39: ...A 1 GHz to 6 GHz Front X1 BiDir Companion device from the base module RF SMA 1 GHz to 6 GHz Front Rx Out Spectrum analyzer test receiver RF SMA 1 GHz to 40 GHz Front RxOUT In Spectrum analyzer test re...

Страница 40: ...ed in the following sections The ambient temperature does not exceed the range specified in the data sheet Signal levels at the input connectors are all within the specified ranges Signal outputs are...

Страница 41: ...S and R S OSP B157WX The operating manual of the R S OSP switch units see chapter on R S EMC32 software The help function of the R S WMS32 software package which is part of the R S RF Test Suite requi...

Страница 42: ...verview extension module 37 D Data sheets 11 Drivers 11 F Frequency ranges 6 Front panel R S OSP B157W8 R S OSP B157W8PLUS 30 R S OSP B157WX 35 G Getting started 9 H Help 10 I Integration 5 Intended u...

Страница 43: ...R S OSP B157W8 R S OSP B157W8PLUS 30 R S OSP B157WX 35 Release notes 11 Removing Existing module R S OSP B157 15 20 RF board R S OSP B157W8 R S OSP B157W8PLUS 31 R S OSP B157WX 36 RF connectors 29 S...

Отзывы: