Remote Control Commands
R&S
®
NRPxxT(N)
87
User Manual 1177.6023.02 ─ 08
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:DATA?
Queries the results of the continuous average result buffer and returns them even if the
buffer is not full.
FETCh<Sensor>[:SCALar][:POWer][:AVG]?
the buffer is full.
Usage:
Query only
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:SIZE
<count>
Sets the size of the result buffer in continuous average mode.
You can enable the buffer using
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:
.
Parameters:
<count>
Range:
1 to 8192
*RST:
1
Example:
BUFF:SIZE 1
See
Chapter 10.2, "Performing a Buffered Continuous Average
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:STATe
<state>
Enables or disables the buffered continuous average mode. If the buffer mode is
enabled, all results generated by trigger events are collected in the sensor until the buf-
fer is filled.
You can set the size of the buffer using
.
Parameters:
<state>
ON | OFF
*RST:
OFF
Example:
BUFF:STAT OFF
[SENSe<Sensor>:][POWer:][AVG:]FAST
<state>
Enables or disables a fast unchopped continuous average measurement. If enabled,
the average count is enforced to 1, and any setting for average count is silently
ignored.
You can increase the measurement accuracy by increasing the duration of the sam-
pling window using
[SENSe<Sensor>:][POWer:][AVG:]APERture
.
See also
"Calculating the measurement time"
Parameters:
<state>
*RST:
OFF
Configuring the Measurement Modes