Remote Control Commands
R&S
®
NRPxxT(N)
70
User Manual 1177.6023.02 ─ 08
Use the command only after the continuous measurement mode has been switched off
(
OFF
).
See also
Chapter 9.5, "Controlling the Measurement"
Example:
See
Chapter 10.2, "Performing a Buffered Continuous Average
Usage:
Event
INITiate:CONTinuous
<state>
Activates/deactivates the continuous measurement mode. In continuous measurement
mode, the power sensor does not go into the idle state after a measurement has been
completed, but immediately executes another measurement cycle.
See also
Parameters:
<state>
ON
Measurements are performed continuously. If a measurement is
completed, the sensor does not return to the idle state but enters
the waiting for trigger state again.
OFF
Ends the continuous measurement mode, and sets the sensor to
the idle state.
*RST:
OFF
Example:
See
Chapter 10.2, "Performing a Buffered Continuous Average
Manual operation:
See
9.5.2
Triggering
In a basic continuous measurement, the measurement is started immediately after the
initiate command, see also
Chapter 9.5.2.2, "Waiting for a Trigger Event"
However, sometimes you want that the measurement starts only if a specific condition
is fulfilled. For example, if a signal level is exceeded, or in certain time intervals. For
these cases, you can define a trigger for the measurement.
Further information:
●
Chapter 9.5.5, "Configuring the Trigger"
9.5.2.1
Trigger States
The power sensor has trigger states to define the exact start and stop time of a mea-
surement and the sequence of a measurement cycle. The following states are defined:
●
Idle
The power sensor performs no measurement. After powered on, the power sensor
is in the idle state.
Controlling the Measurement