63
Specifications
All data valid at 23°C ±5K after 90 minutes warm-up and with 5 3/4 digits
Reading Rates
Function
Setting
Resolution Display
Reading
Rates
(per sec.)
AC Voltage
Slow
5 3/4
400.000
5
Medium
4 3/4
40.000
10
Fast
4 3/4
40.000
200
DC Voltage
Slow
5 3/4
400.000
5
Medium
4 3/4
40.000
10
Fast
4 3/4
40.000
200
AC Current
Slow
5 3/4
200.000
5
Medium
4 3/4
20.000
10
Fast
4 3/4
20.000
200
DC Current
Slow
5 3/4
200.000
5
Medium
4 3/4
20.000
10
Fast
4 3/4
20.000
200
Resistance
Slow
5 3/4
400.000
5
(2-wire)
Medium
4 3/4
40.000
10
Fast
4 3/4
40.000
50
Resistance
Slow
5 3/4
400.000
5
(4-wire)
Medium
4 3/4
40.000
10
Fast
4 3/4
40.000
25
Frequency
Slow
6
999,999
1
Medium
5
99,999
10
Fast
4
9,999
100
Diode
4 3/4
40.000
10
Continuity
4 3/4
40.000
200
Temperature
4
999,9
10
Additional Specifications
DC voltage
Measuring Method
Sigma Delta analog/digital converter
Input Resistance
>10 GΩ (selectable for ranges 400 mV / 4 V
10 MΩ (on all ranges)
CMRR
120 dB for VCM <500 V, 1 kΩ unbalance in the
LO-line and 5 measurements/second
SMRR
>60 dB for 50 or 60 Hz ±0,1% and 5
measurements/second
Input current
60 pA at 25 °C
Overload protection
1000 V for all ranges
AC voltage
Measuring Method
AC coupled True RMS measurement
Input Resistance
1 MΩ parallel <100 pF (on all ranges)
Crest Faktor
Max. 10 (0,5% additional measurement
uncertainty)
CMRR
>60 dB 1 kΩ in the LO-line and frequency
<60 Hz
Overload protection
750 V
rms
(for all ranges)
AC current / DC current
Shunt Resistance
13,75 Ω at 20 mA; 1,25 Ω at 200 mA; 25 mΩ bei
2 A, 10 A
Overload protection
Fuse: F10H250V on the front panel
Resistance
Measuring Method
2-wire and 4-wire
Overload protection
1000 V for all ranges
Continuity
Measuring Method
1 mA constant current
Threshold value
Adjustable in 1Ω steps
Response Time
200 measurements/second
Overload protection
1000 V
Diode Test
Measuring Method
1 mA constant current
Threshold value
Adjustable in 10 mV steps
Response Time
10 measurements/second
Overload protection
1000 V
Temperature
Measuring Method
Resistance measurement with platinum
sensor
Sensor types
PT100, PT500, PT1000
Connection
2-wire, 4-wire
Overload protection
1000 V
Math Functions
Statistics
Min/Max/Average/Standard deviation
Relative measurement
NULL key, offset via keyboard
Logarithmic functions
dB Reference level via keyboard or NULL key
dBm Reference impedance 50/75/600Ω or
freely selectable
Accessories supplied:
Line cord, printed operating manual, R&S®HZ15 silicone
test lead with safety connector and test probe, 1m
(black+red)
Recommended accessories:
R&S®HZ812
PT100 Temperature probe 2-wire
R&S®HZC95
19“ rackmount kit 2RU for R&S®HMC
series
General Specifications
Power supply:
115 V / 230 V ±10%, adjustable voltage selector
on rear panel; Frequency 50Hz / 60Hz
Power consumption:
25 W max., 12 W typ.
Ambient temperature:
0 …+55 °C (operating); –40 …+70 °C (storage)
EMC compliance:
DIN EN 61326-1, DIN EN 55011
Safety compliance:
DIN EN 61010-1, CAN/CSA-C22.2 No.
61010-1-12
Measuring Category:
CAT II, 600 V; CAT I, 1000 V
DC
, 750 V
AC
RMS
Dimensions:
222 x 88 x 280 mm
Weight:
approx. 2,7 kg
Warm-up time:
90 minutes
Data logging
Number of measuring
counts
Internal: 50.000;
external: defined by USB stick capacity.
Rate Log
Min.: 5 ms typ. (in accordance to measuring
function and resolution); Max: 3.600 s
Duration Log
Internal: 250 s...50.000 h;
External: defined by USB stick capacity
Log Daten
Main, 2nd, Time-Stamp
Interfaces
USB 2.0 (TMC and CDC/VCP), Ethernet
10/100 (LXI in Preparation), IEEE-488/GPIB
optional
Programming
SCPI, compatible with Agilent 34401A and
34410A