TECHNISCHE DATEN
Referenztemperatur: 23 °C ±2 °C
TECHNISCHE DATEN
Referenztemperatur: 23 °C ±2 °C
SLIMLINE PROBE HZ355
All data valid at 23 °C after 30 minute warm-up
Electrical specifi cations
Attenuation ratio
1)
:
10:1 ± 2 % at DC
Voltage coeffi cient:
0.0025 %/V (typical)
System bandwidth:
500 MHz (-3 dB)
Probe risetime:
700 ps (10 % - 90 %)(typical)
Maximum rated input voltage:
Measuring category I
2)
:
400 V
rms
(corr. IEC 61010-031)
1250 V
transient overvoltage
Measuring category II
2)
:
300 V
rms
CAT II
(corr. IEC 61010-031)
Note that the max. input voltage rating of
the probe decreases as the frequency of the
applied signal increases.
0
100
200
300
400
0,1
1
10
100
1000
Voltage derating (Measuring category I)
Frequence (MHz)
Amplitude A
C
rm
s
[V] sin
us
Electrical characteristics
Input resistance:
10 MΩ ± 1 %
Input capacitance:
9,5 pF (typical)
Compensation range:
10 pF - 25 pF (typical)
Input coupling of the
measuring instrument:
1 MΩ AC / DC
Note that the input impedance of the probe
decreases as the frequency of the applied
signal increases.
Input impedance
Frequency (MHz)
Mechanical characteristics
Weight (probe only):
48 g
Cable length:
1.3 m
Probe tip Ø :
2.5 mm
Environmental specifi cations
Altitude
operating:
up to 2000 m
non-operating:
up to 15000 m
Temperature range
operating:
0° C to +50° C
non-operating:
-40° C to +70° C
Maximum relative humidity
operating:
80 %
for temperatures up to +30° C
decreasing linearly to 40 % at 50 °C
Pollution degree
3)
: 2
1) Connected to an oscilloscope with an input impedance
of 1 MΩ ± 1 %.
2) Measurement category I
Defi nition: Measurement category I is for measure-
ments performed on circuits not directly connected
to a mains supply.
Examples: Measurements in circuits not derived
from a mains supply and specially protected (internal)
circuits derived from a mains supply. In the latter
case, transient stresses are variable; for that reason
requires that the transient withstand capability of the
equipment is made known to the user.
Measurement category II (CAT II)
Defi nition: Measurement category II is for measure-
ments performed on circuits directly connected to
the low voltage installation.
Examples: Household appliances, portable tools and
similar equipment.
3) Pollution degree 2
Only non conductive POLLUTION. Occasionally, how-
ever, a temporary conductivity caused by conden-
sation must be accepted.
P r o b e H Z 3 5 5
10
100
1k
10k
100k
1M
10M
100M
10
100
1k
10k
100k
1M
10M 100M 1G
|Z
| [
O
h