C o n t r o l s  a n d  r e a d o u t          

7

  Tuning knob (rotary encoder)

Rotary encoder for the parameter entry or change of:
Center frequency CENTER 

3

, SPAN 

4

, START frequency 

5

, STOP 

frequency 

6

, MARKER 

Δ

-MARKER 

17

, REF.LEVEL 

12

, TEST signal 

level 

28

, Intensity INTENS 

9

, FOCUS 

10

, TRACE rotation 

11

, volume 

PHONE 

21

Any entry which would exceed a specifi ed maximum or minimum value 
will not be accepted, the instrument will set the maximum or minimum 
allowable value, a warning will be sounded. 

8

 DISPLAY MODE

Depressing this pushbutton once will dim the character readout 
intensity. The switching sequence is 100%, 50%, 0% and then again 
100. A second depression turns the character readout off. A third one 
will restore the initial setting. The activated interface (RS-232 or USB) 
can be indicated by long pressing (only in connection with the option 
HO720). The interface (RS-232 or USB) can be selected by repeated 
short pressing.

9

 INTENS

The intensity can be varied with the tuning knob 

7

, CW rotation will 

increase, CCW rotation will decrease the intensity. It is recommended 
to increase the intensity only so much as is needed for a well visible 
display, any higher setting will not reveal any more details but defocus 
the trace, also the life of the phosphor in the area of the noise baseline 
will be impaired.

10

 FOCUS

The focus can be adjusted with the tuning knob 

7

. The correct procedu-

re is to fi rst set the intensity and then to adjust the focus for uniformity 
over the whole screen.

11

 TRACE

The trace rotation can be adjusted with the tuning knob 

7

. After de-

pressing this pushbutton a rectangle with a horizontal center line will be 
displayed; this line can be rotated with the tuning knob until it is parallel 
to the graticule center line. In spite of the crt shield provided residual 
infl uences of the earth’s magnetic fi eld may require a readjustment 
when the instrument was moved. A slight barrel or pincushion distortion 
can not be corrected and has no infl uence on the measurements.

12

  dB/div. - dB unit 

(push long)

Short depression:

 alternation of scaling between 10 and 5 dB/div. 

Display in the righthand readout to the right of the attenuator value 
(AT...dB     ... dB/).

Long depression:

 cyclic change of the unit from dBm to dBmV to dBμV. 

The unit selected will be shown in all applicable readout fi elds: (RL ... 
dBm), (ML ... dBm), (TL ... dBm). This pushbutton does not light up.

13

 ATTENUATION 

 

  (0 dB)

These two pushbuttons (which do not light up) switch the input atte-
nuator from (0) 10 to 50 dB in 10 dB steps. Display in the right readout 
fi eld (AT … dB).

STOP

  „0 dB *“ means that this position of the attenuator (no 

attenuation) can only be switched in by an intentional long 
depression of the upper pushbutton. This is provided for 
safety reasons in order to minimize the danger of destruc-
tion of the input stage.

Please note carefully that the maximum specifi ed values for the input 
signal level and a dc content must not be exceeded! This is especially 
important because a spectrum analyzer will not show signals outside its 
specifi ed range (0.1 to 3000 MHz) and also, depending on the settings,  
may display only a portion of the spectrum within its range; excessive 
levels outside the displayed spectrum may cause destruction of the 
input stage.

14

  REF.LEVEL / AUTO 

(push long)

Short depression: adjustment of the reference level either by keyboard 

2

 entry or with the tuning knob 

7

. Prior to any adjustment, the push-

button must be lit, else it must be depressed fi rst. The tuning knob will 
be active immediately; a keyboard entry will be displayed in the last line 
of the left readout fi eld, it will be accepted after a (second) depression 
and displayed in the reference level readout (RL ... dBm), the entry 
display will disappear. Legal values are –110 to +20 dBm. 

AUTO 

 means that the analyzer can be switched to automatic reference 

level selection by a long depression; this will be indicated in the readout 
by (RL * ... dBm). Please note that the 0 dB attenuator position will not 
be used in automatic mode. The AUTO mode is left by another long 
depression. 

Should the noise band already reside at the graticule bottom, the refe-
rence level can not be increased any further, i.e. the noise band can not 
be positioned farther down, a warning will be sounded. The reference 
level can then only be decreased, i.e. the noise band shifted upward; 
this will, however, also decrease the available dynamic range.

The noise band will disappear altogether, if it was already positioned at 
the graticule bottom, if the scaling 

12

 is switched from 10 to 5 dB/div. It 

will become visible again by decreasing the reference level.

Interpretation of measurement results

The measurement results shown in the readout fi elds take all settings 
into account automatically, also the input attenuation; hence all num-
bers displayed represent the true values at the measurement points 
selected in dBm, dBmV, or dBμV, as selected.

STOP

  The reference level refers to the top graticule line, from 

which all measurements are to be derived downward! 
This is exactly the opposite of the procedure with oscillos-
copes. If the reference level is, e.g. RL = 0 dBm, the bot-
tom graticule line corresponds to –80 dBm at 10 dB/div. or 
–40 dBm at 5 dB/div. 

The reference level is equivalent to an offset voltage with scopes, it 
can be selected within the permissible range of –110 to +20 dBm in 
order to facilitate readings; it has no infl uence on the sensitivity or the 
calibration. Quite comparable to a difference amplifi er with offset, the 
dynamic range window of 80 or 40 dB can be shifted within the reference 
level range. 

There are two options for obtaining numerical results: directly from the 
screen or by use of the markers. 

When reading from the screen display, one counts the divisions (cm) 
from the reference level at the top graticule line downward to the point 
of interest and multiplies the divisions with the scaling factor, e.g. 10 
dB/div. If the reference level is, e.g. , RL = 0 dBm, and if the point to be 
measured 1 div (cm) below, its true level will be –10 dBm. 

Much easier is the method of calling the fi rst marker by a short de-
pression of MARKER 

17

 and of positioning it with the tuning knob to 

the point of interest: the level and the frequency can be read directly 
from the readout (ML ... dBm), (MF ... MHz). In practice, mostly the 
levels of spectral peaks are of interest. By a second short depression 
of the MARKER pushbutton the marker will be positioned automatically 
on the highest peak of the spectrum displayed. 
See MARKER 

17

 for a description of the marker functions.

15

 VBW 

(Video bandwidth) 

This pushbutton switches a video fi lter in which reduces the video band-
width from 50 to 4 kHz, indicated in the left readout fi eld (VB 4 kHz). The 
use of this low pass attenuates the noise and increases the visibilty of 
weak signals. This fi lter should not be used with pulsed signals. 

Содержание Hameg HM5530

Страница 1: ...3 G H z S p e c t r u m A n a l y z e r H M 5 5 3 0 Handbuch Manual Manuel Manual Deutsch English Français Español ...

Страница 2: ...erte Anwen dung Die am Messgerät notwendigerweise angeschlossenen Mess und Datenleitungen beeinflussen die Einhaltung der vorgegebenen Grenzwerte in erheblicher Weise Die verwendeten Leitungen sind jedoch je nach Anwendungsbereich unterschiedlich Im praktischen Messbetrieb sind daher in Bezug auf Störaussendung bzw Stör festigkeit folgende Hinweise und Randbedingungen unbedingt zu beachten 1 Daten...

Страница 3: ...hrleistung und Reparatur 7 Wartung 7 Schutzschaltung 7 Netzspannung 7 Kurzbeschreibung der Bedienelemente 8 Test Signal Display 10 Betriebshinweise und Hinweise für erste Messungen 10 Betriebshinweise 10 Erste Messungen 11 Funktionsprinzip des HM5530 11 Normalbetrieb und ZERO SPAN Betrieb 12 Bedienungselemente und Readout 12 RS 232 Interface USB RS 232 Dual Interface Messwertabfrage und Fernsteuer...

Страница 4: ...Hz 120 kHz und 1 MHz YIG Oszillator Pre Compliance EMV Messungen Software für erweiterte Messfunktionen von EMV Messungen im Lieferumfang enthalten RS 232 Schnittstelle optional USB RS 232 für Dokumentation und Steuerung 3 GHz Spektrumanalysator HM5530 HM5530 Erfassung leitungsgebunde ner Störungen 3 GHz Signal mit AM Optional Dual USB RS 232 Schnittstelle HO720 H M 5 5 3 0 ...

Страница 5: ...dbreiten VBW 50 kHz 4 kHz mit automatischer Umschaltung der Sweepzeit 40 80 160 320 und 1000 ms Eingänge Ausgänge Messeingang N Buchse Eingangsimpedanz 50 Ω VSWR ATT 10dB typ 1 5 1 Testsignalausgang N Buchse Ausgangsimpedanz 50 Ω Frequenz 50 MHz 1 kHz Pegel 10 0 dBm in 0 2 dB Stufen Genauigkeit des Pegels 3 dB 0 dBm Versorgungsausgang für Sonden 6 VDC max 100 mA 2 5 mm DIN Klinkenstecker Audioausg...

Страница 6: ...b sicherzustellen muss der Anwender die Hinweise und Warnvermerke beachten die in dieser Bedienungsanleitung im Testplan und in der Service Anleitung enthalten sind Gehäuse Chassis und alle Messanschlüsse sind mit dem Netz schutzleiter verbunden Das Gerät entspricht den Bestimmungen der Schutzklasse I Die berührbaren Metallteile sind gegen die Netzpole mit 2200V Gleichspannung geprüft Das Gerät da...

Страница 7: ...en werden Die Betriebslage ist beliebig Eine ausreichende Luftzirkulation Konvektionskühlung ist jedoch zu gewährleisten Bei Dauerbetrieb ist folglich eine horizontale oder schräge Betriebslage Aufstellbü gel zu bevorzugen Die Betriebslage ist beliebig Eine ausreichende Luftzirkulation Konvektionskühlung ist jedoch zu gewährleisten Bei Dauerbetrieb ist folglich eine horizontale oder schräge Betrie...

Страница 8: ... Langes Drücken Ein und Ausschalten der automatischen Anpassung des Attenuators beim Einstellen des Referenz pegels Funktion eingeschaltet Anzeige R Funktion ausgeschaltet Anzeige RL 15 VBW 15 Video Bandwidth Umschaltung Videofilter zwischen 50 kHz und 4 kHz 16 SELECT 15 Nach einem kurzen Tastendruck auf eine der beiden Tasten wird die gewählte Funktion SAVE oder RECALL und der Speicherplatz für ei...

Страница 9: ...ie maximal zulässigen Eingangs spannungen dürfen nicht überschritten werden Zerstörungs gefahr 30 External TRIGGER 17 BNC Eingang für externes Triggersignal Sweepauslösung 31 ON 17 Taste zum Ein Ausschalten des externen Triggers 32 TEST SIGNAL ON 17 Taste zum Ein Ausschalten des Testsignals 33 OUTPUT 50 Ω 17 Testsignal Ausgangs N Buchse 34 PROBE POWER 17 Stromversorgungsanschluß 6 VDC von Sonden 2...

Страница 10: ...unächst ge prüft werden ob unzulässig hohe Spannungen vorliegen Außerdem ist es empfehlenswert die Messung mit maximaler Abschwächung und dem maximal erfassbaren Frequenzbereich 0 1MHz 3000MHz zu beginnen Trotzdem ist zu berücksichtigen dass unzulässig hohe Signalamplituden auch außerhalb des erfassten Frequenzbereichs vorliegen können die zwar nicht angezeigt werden können z B 3200 MHz jedoch zur...

Страница 11: ...er Referenzpegel REF LEVEL und die Eingangsabschwächung ATT automatisch berücksichtigt Mit einem 2 Marker können Differenzfrequenz und Differenzpegel zum 1 Marker bestimmt werden siehe hierzu die ausführliche Beschreibung Sollen Messwerte ohne Benutzung der Marker erfasst werden so ist zu beachten dass alles vom Bezugswert im Readout RL dBm dies ist der obere Rasterrand her zu rechnen ist Dies ist...

Страница 12: ...Gerät ist also nunmehr ein selektiver Pegelmesser und zeigt den Pegel durch die Lage der Nulllinie mit der gewählten Skalierung logarithmisch an ähnlich einem Oszilloskop das einen DC Pegel linear anzeigt Im Normalbetrieb SPAN 1 bis 3000 MHz wird ein Frequenzbereich angezeigt dessen Umfang von der SPAN Einstellung abhängig ist Beträgt z B die Mittenfrequenz 500 MHz und der Span 1000 MHz Full Span ...

Страница 13: ...rehgeber ist hierdurch sofort aktiviert eine Tastatureingabe wird erst durch einen zweiten Druck auf CENTER 3 übernommen Anzeige links CF Center Frequency Zulässig sind Eingaben von 0 bis 3000 MHz Das der Mittenfrequenzeinstellung ent sprechende Signal wird in Bildschirmmitte angezeigt vorausgesetzt es ist ein von 0 verschiedener Span eingestellt 4 SPAN Span Umfang des auf dem Bildschirm dargestel...

Страница 14: ...telledesSpektrumseinRechteckmithorizon taler Mittellinie Mit dem Drehgeber lässt sich dies um den Mittelpunkt drehen und so einstellen dass sich diese Mittellinie mit der Rastermit tellinie deckt Eine geringe Kissenverzerrung des Rechtecks ist nicht korrigierbar und hat keinen Einfluss auf die Messgenauigkeit 12 dB div dB UNIT Push long KurzesDrücken derTaste Umschaltung der Skalierung von 10 dB Di...

Страница 15: ...s aus dem Messbereich d h außerhalb des Bild schirms fallen 16 SELECT SAVE RECALL Push long Tasten zur Speicherung bzw zum Aufrufen von bis zu 10 Geräteein stellungen es werden nur die 8 im Readout angezeigten Parameter gespeichert DiesegespeichertenWertebleibenauchnachAusschalten erhalten Nach einem Aufruf leuchten wie nach dem Einschalten jedoch nur die Tasten CENTER 3 und WRITE A 26 gleichgülti...

Страница 16: ...ird aus dem vorhandenen Mittelwert und den nächsten Messwerten wird dann erneut ein Mittelwert gebildet und angezeigt Man kann diese gleitende Mittelwertbildung auf dem Bildschirm verfolgen Die Mittelwertbildung verstärkt wiederkehrende Signalteile und unterdrückt stochastische so dass eine starke Rausch verminderung erzielt wird Ist diese Funktion eingeschaltet so ist im Hintergrund auch die MAX ...

Страница 17: ...ertönt nur ein Warnton 28 TEST SIGNAL LEVEL Einstellung der Testsignal Amplitude mit dem Drehgeber 7 von 10 bis 0 dBm in 0 2 dB Stufen 29 INPUT 50 Ω Eingangs N Buchse Ohne Eingangssignal Abschwächung dürfen 10VDC bzw 10dBmnichtüberschrittenwerden MitEingangsabschwä chung 10 bis 50dB sind max 20dBm zulässig Der Außenanschluss der Buchse ist mit dem Gehäuse und damit mit Schutzerde PE verbunden Über...

Страница 18: ...inerAufbau JederBefehl Abfragemussmit 23hex 35dez eingeleitet werden dem 2 folgen Handelt es sich um einen Befehl müssen die Parameter den Buchstaben folgen Abgeschlossen wird jeder Befehl mit der Enter Taste hex 0x0d Es wird nicht zwischen Groß und Kleinschreibweise der Buchstaben unterschieden Die Angabe der Maßeinheit ist immer eindeutig z B Span immer in MHz und wird deshalb nicht mit angegebe...

Страница 19: ...ameterangabe der Centerfrequenz und eine Checksumme der Signalbytes Die Signaldaten belegen folgende Transferdatenbytes trans_byte n sig_data n n 0 bis n 2000 trans_byte 0 sig_data 0 trans_byte 2000 sig_data 2000 Die Checksumme ist ein 24 Bitwert 3 Bytes und wird wie folgt gebildet Checksumme sig_data 0 sig_data 1 sig_data 1999 sig_data 2000 Summe aller Signaldaten Die 24 bit Checksumme belegt fol...

Страница 20: ...r standard For emission the limits for residential commercial and light industry are applied Regarding the immunity susceptibility the limits for industrial environment have been used The measuring and data lines of the instrument have much influence on emission and immunity and therefore on meeting the acceptance limits For different applications the lines and or cables used may be different For m...

Страница 21: ...Warranty and repair 25 Maintenance 25 Protective Switch Off 25 Power supply 25 Front Panel Elements Brief Description 26 Test Signal Display 28 Operational hints 28 First measurements 29 Functional principle of the HM5530 29 Normal operating mode and ZERO SPAN mode 30 Controls and readout 30 RS 232 Interface USB RS 232 Dual Interface Reading measurement results remote control 35 Description 35 USB...

Страница 22: ...olution bandwiths RBW 9 kHz 120kHz and 1MHz YIG oscillator Pre compliance EMI measurements Software for extended measurement functions for EMI measurements included RS 232 Interface optional USB RS 232 for documentation and control 3 GHz Spectrum Analyzer HM5530 HM5530 Measurement of line conducted interference Amplitude modulated 3 GHz signal Dual RS 232 USB Interface Option ...

Страница 23: ...andwidth VBW 50 kHz 4 kHz with automatic selection of sweep time 40 80 160 320 and 1000 ms Inputs Outputs Measuring input N connector Input impedance 50 Ω VSWR ATT 10 dB typ 1 5 1 Testsignal output N connector Output impedance 50 Ω Frequency 50 MHz 1 kHz Level 10 0 dBm in 0 2 dB increments Accuracy of level 3 dB 0 dBm Supply voltage for field probes 6 VDC max 100 mA 2 5 mm DIN jack Audio output Ph...

Страница 24: ...et The instrument operates according to Safety Class I three conductor power cord with protective earthing conductor and a plug with earthing contact The mains line plug must be inserted in a socket outlet provided with a protective earth contact The protective action must not be negated by the use of an extension cord without a protective conductor The mains line plug must be inserted before conn...

Страница 25: ... humidity is up to 80 If condensed water exists in the instrument it should be acclimatized before switching on In some cases e g extremely cold instrument two hours should be allowed before the instrument is put into ope ration The instrument should be kept in a clean and dry room and must not be operated in explosive corrosive dusty or moist envi ronments The instrument can be operated in any po...

Страница 26: ...reference level dependent automatic attenuator setting changes Display in condition ON RL 15 VBW 32 Video bandwidth selection 4 kHz 50 kHz 16 SELECT 33 After briefly pressing either pushbuttons the selected function and the memory location are displayed for a few seconds Within this time the memory locations can be selected by briefly pressing either pushbutton SAVE RECALL If the function SAVE or RE...

Страница 27: ...evels resp voltages must not be exceeded Danger of destruction 30 External TRIGGER 35 BNC input connector for the external trigger signal triggers one sweep 31 ON 35 Toggles the external trigger function on or off 32 TEST SIGNAL ON 35 Toggles the test signal output on or off 33 OUTPUT 50 Ω 35 Output N connector of the test signal 34 PROBE POWER 35 2 5 mm output jack providing 6 VDC 100 mA for field...

Страница 28: ...cy range These would not be displayed but could well overdrive and des truct the input stage Short of destruction any overdrive would create distortions and spurious signals of all sort The frequency range below 100 kHz is not specified any display of spectra in this region may not be reliable The intensity should not be turned up higher than necessary for easy readability any higher setting would ...

Страница 29: ... or 5 dB div At 10 dB div the screen encompasses a dynamic range of 80 dB the bottom graticule line is equivalent to 80 dB if the reference level is e g 0 dB RL 0 dB Functional principle of the HM5530 TheHM5530isaspectrumanalyzerforthefrequencyrangeof100kHzto 3GHz The spectral components of signals in this range can be detected and measured from 110 to 20dBm The signal to be analyzed first passes t...

Страница 30: ...of numerical measurement results Both markers can also be automatically positioned on the peak of the spectrum displayed the deltamarker indicates the difference frequency and the difference amplitude between markers The instrument further features a test signal output reference spec trum which may be used for instrument self tests by connecting it to the input There is also an input for an extern...

Страница 31: ...r to warn that the amplitude measurements may be wrong ZERO SPAN after setting the span to SF 000 000 MHz is a special operating mode The instrument converts to a selective voltmeter of the center frequency signal The level is indicated by the vertical shift of the baseline in 10 or 5 dB div This is similar to a scope displaying a pure dc level 5 START Adjustment display of the start frequency The...

Страница 32: ...rence level either by keyboard 2 entry or with the tuning knob 7 Prior to any adjustment the push button must be lit else it must be depressed first The tuning knob will be active immediately a keyboard entry will be displayed in the last line of the left readout field it will be accepted after a second depression and displayed in the reference level readout RL dBm the entry display will disappear L...

Страница 33: ...splayed a long depression of RECALL will install the setting which is announced by a beep the pushbutton extinguishes the sweep time readout returns If there is no further depression of either SAVE or RECALL after the initial depression the function will be left automatically after 2 s the pushbutton s will extinguish the sweep time display will return In contrast to the HM5014 2 SAVE and RECALL a...

Страница 34: ...ome time and read the result only after no further increase of the level is discernible The function can be disabled by another short depression of the push button MIN HOLD push long This function automatically detects stores and displays the minimum of the averaged spectrum This function is indicated by the lighted pushbutton As is the case for the function MAX HOLD is switched on automatically a...

Страница 35: ...dtothechassisandhencetosafetyground PE the inner conductor is the positive terminal RS 232 Interface USB RS 232 Dual Interface Reading measurement results remote control STOP Attention All terminals of the interface are galvanically connected to the instrument chassis and hence also to the safety ground protective earth PE Elevatedmeasurements i e measurementswheretheinputandoutput terminals and h...

Страница 36: ...xx MHz sr0100 000 E Start frequenz in xxxx xxx MHz st0500 000 E Stop frequenz in xxxx xxx MHz Filter bw1000 E Bandwidth RBW 1000 kHz 120 9 kHz ba1 E Bandwidth automatic ON RBW Auto ba0 E Bandwidth automatic OFF RBW Manual vf0 E Video filter off VBW 50 kHz vf1 E Video filter on VBW 4 kHz Marker mf0500 000 E Marker frequency in xxxx xxx MHz df0100 000 E Delta Marker frequency in xxxx xxx MHz mk0 E all...

Страница 37: ...ata The 24 bit checksum is on the following data bytes trans_byte 2044 1st byte checksum MSB trans_byte 2045 2nd byte checksum trans_byte 2046 3rd byte checksum LSB The center frequency parameter is on the following transmission data bytes trans_byte 2016 C trans_byte 2017 F trans_byte 2018 x trans_byte 2019 x trans_byte 2020 x trans_byte 2021 x trans_byte 2022 trans_byte 2023 x trans_byte 2024 x ...

Страница 38: ...mation générale concernant le marquage CE Les instruments HAMEG répondent aux normes de la directive CEM Le test de conformité fait par HAMEG répond aux normes génériques actuelles et aux normes des produits Lorsque différentes valeurs limites sont applicables HAMEG applique la norme la plus sévère Pour l émission les limites concernant l environnement domestique commercial et industriel léger son...

Страница 39: ...cation 43 Conditions ambiantes 43 Garantie et réparation 43 Entretien 43 Circuit de protection 43 Tension du réseau 43 Description sommaire des éléments de commande 44 Consignes d utilisation 46 Premières mesures 47 Principe de fonctionnement du HM5530 47 Fonctionnement normal et fonctionnement avec une excursion nulle 48 Éléments de commande et Readout 50 Interface RS 232 Commande à distance 54 D...

Страница 40: ...ante de résolution RBW 9 kHz 120 kHz et 1 MHz Oscillateur YIG Mesure de pré qualification CEM Logiciel pour les fonctions de mesures élargies de CEM inclus Interface RS 232 en option USB RS 232 pour transfert des données et commande Analyseur de spectre 3 GHz HM5530 HM5530 Capture de signaux parasites Signal 3 GHz Modulé en amplitude Interface double RS 232 USB HO720 en option H M 5 5 3 0 ...

Страница 41: ...ge 40 80 160 320 et 1000 ms Entrées Sorties Entrée de mesure Connecteur N Impédance d entrée 50 Ω VSWR ATT 10 dB typique 1 5 1 Sortie signal de test Connecteur N Impédance de sortie 50 Ω Fréquence 50 MHz 1kHz Niveau 10 0 dBm par pas de 0 2 dB Précision du niveau 3 dB 0 dBm Alimentation pour sondes de champ 6 VDC max 100 mA prise jack 2 5 mm Sortie audio Phone prise jack 3 5 mm Interface RS 232 con...

Страница 42: ...conforme aux dispositions de la classe de protection I L isolement entre les parties métalliques accessibles et les bornes du secteur a été contrôlé avec une tension continue de 2200 V Pour des raisons de sécurité l oscilloscope doit uniquement être branché à des prises avec terre conformes à la réglementation Il faut brancher la fiche secteur avant la connexion des circuits de mesure Il est interd...

Страница 43: ...réparation Les instruments HAMEG sont soumis à un contrôle qualité très sévère Chaque appareil subit un test burn in de 10 heures avant de quitter la production lequel permet de détecter pratiquement chaque panne prématurée lors d un fonctionnement intermittent L appareil est en suite soumis à un essai de fonctionnement et de qualité approfondi au cours duquel sont contrôlés tous les modes de fonc...

Страница 44: ...désactivationdel adaptationauto matique de l atténuateur lors du réglage du niveau de référence fonction activée affichage de R fonction désactivée affichage de RL 15 VBW 52 Bande passante vidéo sélection du filtre vidéo 50 kHz ou 4 kHz 16 SELECT 53 Une brève pression sur l une des deux touches affiche pendant quelques secondes la fonction sélectionnée SAVE ou RECALL ainsi que l emplacement de la mémo...

Страница 45: ...0 Ω 55 Prise N d entrée Il ne faut pas dépasser les tensions d entrée maximale admissibles risque de destruction 30 External TRIGGER 55 Entrée BNC pour un signal de déclenchement externe déclenchement de la wobulation 31 ON 55 Touche d activation désactivation du déclenchement externe 32 TEST SIGNAL ON 55 Touche d activation désactivation du signal de test 33 OUTPUT 50 Ω 55 Sortie du signal de tes...

Страница 46: ...teur de tension d entrée HZ560 afin d éviter tout risque de destruction de l atténuateur d entrée et ou du premier étage mélangeur Lorsdel analysedesignauxinconnus ilconvienttoutd aborddevérifier s il existe des tensions excessivement élevées Il est en outre recom mandé de commencer la mesure avec une atténuation maximale et à la plage de fréquences maximale détectable 0 1 MHz 3000 MHz Il faut malgr...

Страница 47: ...ence et du niveau à l endroit du curseur Le niveau de référence REF LEVEL et l atténuation d entrée ATTN sont automatiquement pris en compte lors de l affichage du niveau Si vous voulez relever une valeur sans utiliser le curseur commencez par déterminer l écart en dB entre la ligne supérieure de la grille qui correspondauniveauderéférenceaffichéparleReadout RL dBm et la crête du signal N oubliez pa...

Страница 48: ... oscillateur local est de 3537 4 MHz ce qui donne une différence de 3537 3 MHz 3537 4 MHz 100 kHz Pour 1000 MHz 4537 3 MHz 1000 MHz la différence est de nouveau de 3537 3 MHz Après le 1er étage mélangeur les signaux décrits précédemment sont acheminés à un filtre passe bande filtre FI dont la fréquence centrale est de 3537 3 MHz Seuls peuvent ainsi parvenir à la sortie du filtre passe bande la différ...

Страница 49: ... balayage soit terminé et que le prochain commence L appareil permet aussi de sélectionner directement une fréquence de départ et d arrêt Les données mémorisées du signal peuvent être conditionnées et transmises par le biais de l interface série à un PC qui permet égale ment de commander l appareil à distance Les fonctions disponibles sont notamment le calcul de la moyenne la détection de la valeu...

Страница 50: ...sur la touche de fonction allumée valide la valeur saisie et la transfère dans le champ correspondant du Rea dout Une saisie suivie d une pression sur une touche non allumée est ignorée et effacée Si la valeur saisie est supérieure à une limite autorisée seule la valeur maximale sera alors adoptée et aucun signal sonore d alerte ne sera émis 3 CENTER Réglage de la fréquence centrale avec le CLAVIE...

Страница 51: ...de la luminosité avec le codeur ROTATIF 7 La luminosité augmente en tournant vers la droite diminue en tournant vers la gauche Il ne faut augmenter la luminosité qu à un niveau qui permet une lecture correcte car une luminosité excessive n apporte aucune amélioration et dégrade l astigmatisme 10 FOCUS Réglage de l astigmatisme avec le CODEUR ROTATIF 7 Le réglage correct produit un astigmatisme rég...

Страница 52: ... bord inférieur de la grille correspond alors à 80 dBm à 10 dB div ou à 40 dBm à 5 dB div Le niveau de référence correspond à une tension d offset sur un oscilloscope il peut être réglé à une valeur quelconque au sein de la plage autorisée en vue de faciliter la lecture mais il n influence pas la sensibilité ni le calibrage Comme dans le cas d un oscilloscope avec amplificateurdifférentielettensiond...

Страница 53: ... Une pression prolongée active une sélection automatique de la bande passante de résolution optimale ce qui est indiqué dans le Readout par R kHz Une nouvelle pression prolongée désactive la fonction La bande passante est encore réduite si le filtre vidéo VBW 15 a été activé VB4kHz Lesignaldemesuredéclenchelefiltredemanièreàce que soit représentée wobulée la courbe de passage du filtre à chaque fois ...

Страница 54: ...ple lors des analyses pré liminaires de CEM En raccordant une antenne à l entrée de l analyseur de spectre et en sélectionnant une excursion nulle avec SPAN 4 SF 000 000 MHz le réglage de la fréquence centrale 3 permet alors d effectuer un accord sur un émetteur Il faut notamment tenir compte ici des dispositions légales nationales 23 VIEW B Une pression sur cette touche ne l allume que si un spec...

Страница 55: ... Cette interface bidirectionnelle permet de commander l appareil de mesure ou de collecter des paramètres de réglage ou des informations sur le signal depuis un ordinateur Câble RS 232 Le câble doit avoir moins de 3 m de long être blindé et ne doit pas être croisé connexion directe 1 1 Le brochage de l interface RS 232 prise femelle Sub D 9 broches est le suivant Broche 2 Tx Data transmission des ...

Страница 56: ...ency CFxxxx xxx in MHz sp E Span frequency SPxxxx xxx in MHz sr E Start frequency SRxxxx xxx in MHz st E Stop frequency STxxxx xxx in MHz Marker mf E Marker frequency MFxxxx xxx in MHz df E Delta frequenz DFxxxx xxx in MHz mk E Marker mode MKx x 0 OFF x 1 Marker1 x 2 M1 2 lv E aktiv Marker level ML xxx x in dB Unit MK1 or aktiv Delta Level DL xxx x in dB MK2 Test signal tl E Test signal level TL x...

Страница 57: ...onnées du signal sont le résultat de 2001 conversions analogique numérique pendant une période de vobulation Position X Lepremieroctet sig_data 0 correspondaupremierpointsurl écran cathodique celui qui coïncide avec la ligne gauche de la grille Il est suivi de manière linéaire par tous les autres octets jusqu à sig_dat 2000 lequel coïncide avec la ligne droite de la grille La fréquence de chacun d...

Страница 58: ...parael campodelosnegocioseindustrias asícomoeldelaspequeñasempresas clase 1B En relación a los márgenes de protección a la perturbación externasehanelegidolosvaloreslímiteválidosparalaindustria Loscables o conexiones conductores acoplados necesariamente a un osciloscopio para la transmisión de señales o datos influyen en un grado elevado en el cumplimiento de los valores límite predeterminados Los ...

Страница 59: ...paraciones 63 Mantenimiento 63 Desconexión de seguridad 63 Tensión de red 63 Cambio de fusible de red del equipo 63 Descripción abreviada de los elementos de mando 64 Presentación de la señal de test 66 Indicaciones de funcionamiento 66 para primeras mediciones 66 Primeras mediciones 67 Principio de funcionamiento del HM5530 67 Modo de funcionamiento Normal y modo ZERO SPAN 68 Mandos de Control y ...

Страница 60: ...ncho de banda RBW de 9kHz 120kHz y 1MHz Oscilador YIG Medidas de pre homologaciones EMC Se adjunta el software que contiene funciones ampliadas de medida para mediciones EMC Interfaz RS 232 opcional USB RS 232 para documentación y control Analizador de Espectros de 3 GHz HM5530 HM5530 Captura de ruidos dependientes de la potencia Señal de 3 GHz con AM Opcional Interfaz DUAL RS 232 USB HO720 H M 5 ...

Страница 61: ... ms 80 ms 160 ms 320 ms y 1000ms Entradas Salidas Entrada de medida Borne N Impedancia de entrada 50 Ω VSWR Attn 10 dB tip 1 5 1 Salida de señal de test Borne N Impedancia de salida 50 Ω Frecuencia 50 MHz 1 kHz Nivel 10 0 dBm en pasos de 0 2 dB Exactitud del nivel 3 dB 0 dBm Alimentación para sondas 6 VDC máx 100 mA Conector DIN tipo banana de 2 5 mm Salida de audio Phone Conector banana de 3 5 mm...

Страница 62: ...ado de seguridad del aparato y garantizar un manejo seguro La caja el chasis y todas las conexiones de medida están conectadas al contacto protector de red tierra El aparato corresponde a la clase de protección I Las partes metálicas accesibles para el usuario están comprobadas con respecto a los polos de red con 2200V Por razones de seguridad el aparato sin transformador de aislamiento solamente ...

Страница 63: ...medida HAMEG ha sido fabricado con la máxima diligencia yhasidocomprobadoantesdesuentregapornuestrodepartamentode controldecalidad pasandoporunacomprobacióndefatigaintermitente de 10 horas A continuación se han controlado en un test intensivo de calidad todas las funciones y los datos técnicos Son válidas las normas de garantía del país en el que se adquirió el producto de HAMEG Por favor contacte...

Страница 64: ...as se corresponden con las des cripcio nes explícitas bajo el capítulo Mandos de control y readout 14 REF LEVEL 71 Pulsación breve Ajuste del nivel de referencia mediante el tecla do 2 o mediante el mando giratorio 7 indicación RL o R AUTO Pulsación prolongada Activa o desactiva la adaptación automática del atenuador al ajustar el nivel de referencia Función activada Indicación R Función desactiva...

Страница 65: ...eñal actual A a la memoria de referencia B 28 TEST SIGNAL LEVEL 74 Ajuste del nivel de la señal de test mediante el teclado 2 o mediante el mando giratorio 7 indicación TL 29 INPUT 50 Ω 74 Borne de entrada de tipo N Las tensiones de entrada máximas permitidas no deben superarse hay peligro de dañar el equipo 30 EXTERNAL TRIGGER 74 Entrada tipo BNC para una señal externa de disparo Iniciación del b...

Страница 66: ...e Si se utiliza un reproductor de redes LISN se debe proteger la entra da del analizador de espectros mediante un limitador de transientes HZ560 De lo contrario se corre el riesgo de deteriorar el atenuador de entrada y o la primera etapa del mezclador Antes de examinar señales sin identificar tiene que verificarse la presencia de tensiones altas inaceptables También se reco mienda empezar la medi...

Страница 67: ...s valores de la marca MF ML En la indicación de los valores en pantalla se tienen en cuenta de forma automática los valores de referencia REF LEVEL y de atenuación de entrada ATT Con la segunda marca disponible se pueden determinar las diferencias en frecuencia y en nivel con la primera marca ver las descripciones específicas Si se desean cuantificar los valores de medida sin la ayuda de las marcas ...

Страница 68: ...recen cuando el tiempo de medida es demasiado corto o la velocidad de barrido es demasiado rápida y no cumple con los requisitos preestablecidos por el filtro de la frecuencia intermedia y o del filtro de vídeo en lo que que corresponde a tiempo de oscilación El equipo muestra entonces en el campo del tiempo de medida en pantalla SW uncal Modo de funcionamiento Normal y modo ZERO SPAN Al medir se di...

Страница 69: ...t en el margen izquierdo y derecho superior y al mismo tiempo se presenta sin una señal acoplada a la entrada en el margen inferior de la pantalla la línea de base como una banda de ruidos más o menos ancha Indicación Al apagar el equipo se pierden todos los contenidos de la memoria con excepción de las memorias que se utilizan para memorizar los ajustes de los mandos Al encender el equipo se ajus...

Страница 70: ...e inicio actual Si se prueba de ajustar una combinación sin sentido es decir una fre cuencia de inicio mayor a la frecuencia de paro el equipo ajusta ambas con el mismo valor y se conmuta a ZERO SPAN ver bajo SPAN 6 STOP Ajuste de la frecuencia de paro Primero se deberá pulsar la tecla para que ésta se ilumine El mando rotativo 7 queda entonces activado una introducción numérica por teclado 2 serí...

Страница 71: ...ucir el nivel de referencia la banda de ruido posicionarse nuevamente en zona visible Evaluación de la medida Losvaloresdemedidapresentadostienenencuentadeformaautomáti ca todos los ajustes realizados por lo tanto también la atenuación de entrada ajustada y presentan la amplitud de la señal real de los puntos de medida seleccionados en dBm dBmV o dBμV STOP El nivel de referencia se refiere al borde...

Страница 72: ...terizada por un rombo y el readout presenta entonces en vez de MF y ML con signos correctos la diferencia en frecuencia DF Delta Frequency y la diferencia en nivel DL Delta Level entre ambas marcas Una segunda pulsación prolongada po siciona la marca en forma de rombo sobre el máximo del espectro presentado La diferencia en frecuencia sólo puede ser variada con el mando giratorio Estando ambas mar...

Страница 73: ... tecla se conmuta de MIN HOLD a MAX HOLD Mediante una pulsación más prolongada se conmuta de MAXHOLDaMINHOLD Lafuncióncalculaautomáticamentelosniveles de la señal capturados por el equipo sólo se actualizará la presentaci ón si aparece un valor inferior al actual De esta forma se posibilita la medición fidedigna del nivel mínimo incluso cuando aparecen breves variaciones de señal o interrupciones d...

Страница 74: ...n ello a masa detierra PE Traspasarloslímitesseñaladospuededeteriorarlaetapa de entrada del analizador de espectros 30 EXTERNAL TRIGGER Borne BNC para el acoplamiento de una señal de disparo exterior Nivel bajo LOW 0 0 8V nivel alto HIGH 2 5V 5 0V con disparo sobre la pendiente positiva umbral de disparo típico 1 3V tensión máxima de entrada 10V 31 ON Tecla para activar el disparo externo 32 TEST ...

Страница 75: ... trata de una orden deberán seguir los parámetros a las letras Cada órden se finaliza con la tecla Enter hex 0x0d No se diferencia entre letras mayúsculas o minúsculas La indicación de la unidad siempre es clara p ej Span siempre en MHz y por esa razón no se introducirá Lista de las órdenes de ajuste E significa tecla Enter signo cr 0x0D carriage return retroceso del carro kl0 E Key Lock off control...

Страница 76: ...alinealhastasig_dat 2000 quecoincide entonces con la línea derecha de la retícula La frecuencia de los puntos individuales se puede determinar de la frecuencia central y del Span Frecuencia x Frecuencia central 0 5 x Span Span x x 2000 X 0 2000 Posición del punto sig_data x Posición Y el octavo valor hex 00 hasta FF de cada célula de memoria de sig_data x tiene la siguiente referencia con la señal...

Страница 77: ...s 0 4 dB con 10dB Div y 0 2 dB con 5dB Div El nivel de un punto y se puede calcular Para y 229 posición del nivel de referencia Level en dBm y Nivel de referencia dBm 229 y x 0 4 dB con 10dB Div Para y 229 posición del nivel de referencia Level en dBm y Nivel de referencia dBm y 229 x 0 4 dB con 10dB Div I n t e r f a z R S 2 3 2 c o n t r o l r e m o t o ...

Страница 78: ...78 Reservado el derecho de modificación ...

Страница 79: ...79 Reservado el derecho de modificación ...

Страница 80: ...riestraße 6 A Rohde Schwarz Company D 63533 Mainhausen registriered trademark Tel 49 0 61 82 800 0 DQS Certification DIN EN ISO 9001 2000 Fax 49 0 61 82 800 100 Reg Nr 071040 QM sales hameg de 42 5530 0040 authorized dealer Oscilloscopes Spectrum Analyzer Power Supplies Modular System Series 8000 Programmable Instruments Series 8100 ...

Отзывы: