Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
106
User Manual 1173.9286.02 ─ 15
8.9 Using Limit Lines
The following commands are necessary to set up and configure limit lines.
............................................................................. 106
................................................................................... 109
Creating and Editing Standard Limit Lines
............................................................113
8.9.1 Using Phase Noise Limit Lines
.........................................................................................106
CALCulate:PNLimit:FC1 <Frequency>
............................................................................. 106
CALCulate:PNLimit:FC2 <Frequency>
............................................................................. 106
CALCulate:PNLimit:FC3 <Frequency>
............................................................................. 106
CALCulate:PNLimit:FC4 <Frequency>
............................................................................. 106
..............................................................................................107
.............................................................................................107
............................................................................................... 107
...............................................................................................108
CALCulate:PNLimit:SLOPe<segment>
.............................................................................108
CALCulate:PNLimit:COPY<k>
Creates a new user limit line from the data of a phase noise limit line.
Suffix:
<k>
.
1...8
Number of the limit line the phase noise limit line is copied to. An
existing limit line in that slot is overwritten.
Example:
CALC:PNL:COPY2
Copies the phase noise limit line to limit line 3.
Usage:
Event
Manual operation:
See
CALCulate:PNLimit:FC1 <Frequency>
CALCulate:PNLimit:FC2 <Frequency>
CALCulate:PNLimit:FC3 <Frequency>
CALCulate:PNLimit:FC4 <Frequency>
CALCulate:PNLimit:FC5
<Frequency>
This command defines the start frequency of a segment of a phase noise limit line.
CALCulate:PNLimit:FC1 is available for special phase noise limit lines with one seg-
ment or more.
CALCulate:PNLimit:FC2 is available for special phase noise limit lines with two seg-
ment or more.