Remote Control
R&S
®
ESRP
869
User Manual 1175.7445.02 ─ 08
.........................................................................................870
............................................................................................. 872
................................................................................... 873
SOURce<n>:POWer[:LEVel][:IMMediate][:AMPLitude]
.......................................................873
SOURce<n>:POWer[:LEVel][:IMMediate]:OFFSet
............................................................. 873
............................................................................................874
............................................................................................ 874
.............................................................................................875
OUTPut[:STATe]
<State>
This command switches the selected tracking generator on or off.
The command is available with a tracking generator (option R&S
FSV-B9).
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
OUTP1 ON
Switches on the external tracking generator 1.
Usage:
SCPI confirmed
Manual operation:
See
"Source RF Internal (On/ Off)"
[SENSe:]CORRection:COLLect[:ACQuire]
<MeasType>
This command initiates a reference measurement (calibration). The reference mea-
surement is the basis for the measurement normalization.
To obtain a correct reference measurement, a complete sweep with synchronization to
the end of the sweep must have been carried out. This is only possible in the single
sweep mode.
The command is available if a tracking generator is installed and has been turned on.
Parameters:
<MeasType>
THRough
"TRANsmission" mode: calibration with direct connection
between tracking generator and device input
"REFLection" mode: calibration with short circuit at the input
OPEN
only allowed in "REFLection" mode: calibration with open input
Remote Commands to Control the Tracking Generator