Remote Control
R&S
®
ESRP
868
User Manual 1175.7445.02 ─ 08
SCPI command:
TRACE:IQ:EGATE:TYPE EDGE
, see
●
Level triggered recording
After a trigger signal, all data is recorded in which the gate signal is set to 1, which
means it has exceeded a level. In this case, the gate signal can be generated by
the IFP trigger, for example: each time the IFP level is exceeded, the IFP trigger
signal is set to 1 and the samples in this area are recorded as gate samples.
The number of complex samples to be recorded prior to the trigger event can be
selected (see
on page 858) for all available trigger sources,
except for "Free Run".
SCPI command:
TRACE:IQ:EGATE:TYPE LEV
, see
11.7
Remote Commands to Control the Tracking Generator
The following commands control the optional tracking generator.
Controlling the Internal Tracking Generator
.......................................................... 868
Controlling the External Tracking Generator
.........................................................875
11.7.1
Controlling the Internal Tracking Generator
[SENSe:]CORRection:COLLect[:ACQuire]
........................................................................869
....................................................................................... 870
..........................................................................................870
Remote Commands to Control the Tracking Generator