Remote Control
R&S
®
ESRP
854
User Manual 1175.7445.02 ─ 08
Example:
CALC:MARK:FUNC:BPOW:STAT ON
Manual operation:
See
CALCulate<n>:DELTamarker<m>:FUNCtion:BPOWer:MODE
<Mode>
This command selects the way the results for a band power delta marker are dis-
played.
Suffix:
<n>
.
irrelevant
<m>
irrelevant
Parameters:
<Mode>
POWer
Result is displayed as a power in dBm.
DENSity
Result is displayed as a density in dBm/Hz.
*RST:
POW
Example:
CALC:DELT:BPOW:MODE DENS
CALCulate<n>:DELTamarker<m>:FUNCtion:BPOWer:RESult?
This command queries the results of the band power measurement.
Suffix:
<n>
.
irrelevant
<m>
irrelevant
Return values:
<Power>
Signal power over the delta marker bandwidth.
Example:
Activate the delta band power marker:
CALC:DELT:BPOW:STAT ON
Select the density mode for the result:
CALC:DELT:BPOW:MODE DENS
Query the result:
CALC:DELT:BPOW:RES?
Response:
20dBm/Hz
Usage:
Query only
CALCulate<n>:DELTamarker<m>:FUNCtion:BPOWer:SPAN
<Span>
This command defines the bandwidth around the delta marker position. The span is
indicated by lines in the diagram.
Suffix:
<n>
.
irrelevant
Remote Commands in Spectrum Analyzer Mode