Spectrum Measurements
R&S
®
ESRP
265
User Manual 1175.7445.02 ─ 08
...................................................................... 269
................................................................................ 270
.............................................................................. 270
.................................................................................................270
............................................................................................ 270
Spurious Emissions
Opens a submenu to configure the Spurious Emissions measurement and immediately
starts a measurement.
For general information on performing Spurious Emissions measurements see
ter 6.1.1.6, "Measuring Spurious Emissions"
Remote command:
Sweep List ← Spurious Emissions
Opens a submenu to edit the sweep list and displays the "Sweep List" dialog box.
Note:
If you edit the sweep list, always follow the rules described in
"Measuring Spurious Emissions"
Sweep List dialog box ← Sweep List ← Spurious Emissions
After a preset, the sweep list contains a set of default ranges and parameters. For
each range, you can change the parameters listed below. To insert or delete ranges,
use the "Insert Before Range", "Insert After Range", "Delete Range" softkeys. The
measurement results are not updated during editing but on closing the dialog box.
For details and limitations regarding spurious emissions configuration, see
ter 6.1.1.6, "Measuring Spurious Emissions"
Range Start / Range Stop ← Sweep List dialog box ← Sweep List ← Spurious
Emissions
Sets the start frequency/stop frequency of the selected range. Follow the rules descri-
bed in
Chapter 6.1.1.6, "Measuring Spurious Emissions"
In order to change the start/stop frequency of the first/last range, select the appropriate
span with the SPAN key. If you set a span that is smaller than the overall span of the
ranges, the measurement includes only the ranges that lie within the defined span and
have a minimum span of 20 Hz. The first and last ranges are adapted to the given
span as long as the minimum span of 20 Hz is not violated.
Frequency values for each range have to be defined relative to the center frequency.
The reference range has to be centered on the center frequency. The minimum span of
the reference range is given by the current TX Bandwidth.
Remote command:
[SENSe:]LIST:RANGe<range>[:FREQuency]:STARt
[SENSe:]LIST:RANGe<range>[:FREQuency]:STOP
Filter Type ← Sweep List dialog box ← Sweep List ← Spurious Emissions
Sets the filter type for this range. For details on filters see also
"Selecting the Appropriate Filter Type"
Measurements