Spectrum Measurements
R&S
®
ESRP
259
User Manual 1175.7445.02 ─ 08
Example
Figure 6-6: Sweep list using Fast SEM mode
In
, a sweep list is shown for which Fast SEM is activated. The formerly 5
separately defined ranges are combined to 2 sweep ranges internally.
6.1.1.6
Measuring Spurious Emissions
The Spurious Emissions measurement defines a measurement that monitors unwan-
ted RF products outside the assigned frequency band generated by an amplifier. The
spurious emissions are usually measured across a wide frequency range. The Spuri-
ous Emissions measurement allows a flexible definition of all parameters. A result table
indicates the largest deviations of the absolute power from the limit line for each range,
and the results can be checked against defined limits automatically.
Measurements