Remote Control
R&S
®
ESRP
721
User Manual 1175.7445.02 ─ 08
Suffix:
<range>
.
1...20
range
Parameters:
<Type>
NORMal
Gaussian filters
CFILter
channel filters
RRC
RRC filters
P5
5 Pole filters
*RST:
NORM
The available bandwidths of the filters are specified in the data
sheet.
Example:
LIST:RANG1:FILT:TYPE RRC
Sets the RRC filter type for range 1.
Manual operation:
See
[SENSe:]LIST:RANGe<range>[:FREQuency]:STARt
<Frequency>
This command sets the start frequency for the specified range.
In order to change the start/stop frequency of the first/last range, select the appropriate
span. If you set a span that is smaller than the overall span of the ranges, the mea-
surement includes only the ranges that lie within the defined span and have a minimum
span of 20 Hz. The first and last range are adapted to the given span as long as the
minimum span of 20 Hz is not violated.
Suffix:
<range>
.
1...20
range
Parameters:
<Frequency>
see rules in chapter 4, "Ranges and settings"
*RST:
-250.0 MHz (range 1), -2.52 MHz (range 2), 2.52
MHz (range 3)
Example:
LIST:RANG1:STAR 100000000
Sets the start frequency for range 1 to 100 MHz.
Manual operation:
See
[SENSe:]LIST:RANGe<range>[:FREQuency]:STOP
<Frequency>
This command sets the stop frequency for the specified range. See also
]LIST:RANGe<range>[:FREQuency]:STARt
Remote Commands in Spectrum Analyzer Mode