Remote Control
R&S
®
ESRP
729
User Manual 1175.7445.02 ─ 08
Parameters:
<FileName>
Example:
ESP:PRES:STOR
'WCDMA\3GPP\DL\PowerClass_31_39.xml'
Stores the settins in the
PowerClass_31_39.xml
file in the
C:
\R_S\instr\sem_std\WCDMA\3GPP\DL
directory.
Manual operation:
See
Controlling the Measurement
The following commands control the measurement itself.
INITiate<n>:ESPectrum
This command starts a Spectrum Emission Mask measurement.
Suffix:
<n>
.
irrelevant
Example:
INIT:ESP
Starts a Spectrum Emission Mask measurement.
Manual operation:
See
[SENSe:]SWEep:MODE
<Mode>
This command selects the spurious emission and spectrum emission mask measure-
ments.
Parameters:
<Mode>
AUTO | ESPectrum | LIST
AUTO
Switches to "Spectrum" measurement mode or stays in the cur-
rent mode if it is not ESP/LIST
ESPectrum
"Spectrum Emission Mask" measurement mode
LIST
"Spurious Emissions" measurement mode
*RST:
AUTO
Example:
SWE:MODE ESP
Sets the Spectrum Emission Mask measurement mode.
Manual operation:
See
See
See
See
Remote Commands in Spectrum Analyzer Mode