Receiver Mode
R&S
®
ESRP
160
User Manual 1175.7445.02 ─ 08
Input 50 Ω/75 Ω
Uses 50 Ω or 75 Ω as reference impedance for the measured levels. Default setting is
50 Ω.
The setting 75 Ω should be selected if the 50 Ω input impedance is transformed to a
higher impedance using a 75 Ω adapter of the RAZ type (= 25 Ω in series to the input
impedance of the instrument). The correction value in this case is 1.76 dB = 10 log (75
Ω/50 Ω).
All levels specified in this Operating Manual refer to the default setting of the instru-
ment (50 Ω).
Remote command:
5.3.3
Selecting the Bandwidth
The bandwith settings are available in the [BW] menu.
..............................................................................................160
Res BW Manual
Opens an input field to define the measurement or resolution bandwidth.
The R&S
ESRP supports a selected set of resolution bandwidths. If you enter a num-
ber that is not supported, the R&S
ESRP rounds the value up to next available band-
width.
You can also select some bandwidths directly with the corresponding softkeys in the
"Bandwidth" menu.
For more information see
Chapter 5.2.1, "Measurement Bandwidth"
Remote command:
[SENSe:]BANDwidth|BWIDth[:RESolution]
CISPR RBW Uncoupled
Cancels the coupling of the IF bandwidth to the frequency range with the activated
quasipeak detector, CISPR Average or RMS average detector.
See also
Remote command:
[SENSe:]BANDwidth|BWIDth[:RESolution]:AUTO
Filter Type
Selects the filter type.
The available resolution bandwidths depend on the filter selection.
The R&S
ESRP provides the following filter types:
●
EMI CISPR / MIL
Common Measurement Settings