<CellParameterID>
Cell parameter ID, identifying the cell and determining the scram-
bling code.
Range:
0 to 127
*RST:
0
<MidambleK>
US2 | US4 | US6 | US8 | US10 | US12 | US14 | US16
Maximum number of users for one timeslot (2 to 16 users). The
value has an impact on the used midamble sequences.
*RST:
US16
<MTYPe>
QPSK | EPSK | QAM
Modulation type
QPSK
: quadrature phase shift keying
EPSK
: eight phase shift keying (8PSK)
QAM
: quadrature amplitude modulation (16-QAM)
*RST:
QPSK
<ActiveSlots>
Bit pattern indicating the active uplink slots to be measured
0 = inactive / 1 = active, leading zeros may be omitted
From right to left, the bits correspond to the first to sixth uplink slot,
relative to the trigger event
Example 1:
Slot 3 and slot 5 are active, but only slot 5 shall be measured. An
IF trigger is used, so the trigger event occurs before slot 3.
Correct bit pattern: #B000100 (or #B100)
Example 2:
Same signal (slot 5 shall be measured), but an external trigger or
waveform marker is used, so the trigger event is expected before
slot 0.
Correct bit pattern: #B010000 (or #B10000)
Range:
#B000000 to #B111111
*RST:
#B1
<Level>
Expected nominal power in the segment. The range of the expec-
ted nominal power can be calculated as follows:
Range (Expected Nominal Power) = Range (Input Power) + Exter-
nal Attenuation - User Margin
Range:
–47 dBm to 35 dBm for the input power at the RF
COM connectors (please notice also the ranges quo-
ted in the data sheet).
*RST:
0 dBm
Default unit: dBm
<Frequency>
Center frequency of the RF analyzer
Range:
100E+6 Hz to 6E+9 Hz
*RST:
2.0108E+9 Hz
Default unit: Hz