CMU-Z6
Operating Instructions
1150.0199.12
5
E-3
The CMU states that the CMU-Z6 must be connected and that Mode B must be selected as
described above.
Press
ENTER
to continue the modulation correction.
The progress of the correction is monitored in the
Status
and
Progress
fields. Once the correction is
terminated, the
Test
status indication changes to
HLT
and the CMU is ready to perform an improved
wideband modulation analysis.
Important Note:
A warm-up time of at least 30 minutes is necessary to reach the normal operating temperature
of the CMU; otherwise the correction values might be impaired.
Checking the Corrected Measurement
The accuracy of the corrected modulation measurement can be checked by using a standard QPSK or
WCDMA signal from a signal generator. We propose the following equipment and test setup:
Equipment
Properties
Appropriate R&S device
Stock no.
Use
SMIQ03
1084.8004.03
QPSK
Signal generator
300 kHz to 3.3 GHz
QPSK / WCDMA
SMIQ03B
with option SMIQ-B47
1125.5555.03
WCDMA
Coax cable
N to N, approx. 1 m length
RF connection cable
0100.7670.10
RF connection
SMIQ –> CMU
Coax cable
BNC to BNC (50
Ω
),
approx. 1.8 m length
RF connection cable
0100.6945.20
Synchronize CMU to 10 MHz
reference signal from SMIQ