The CMU states that the CMU-Z6 must be connected and that Mode B must be selected as
described above.

 Press 

ENTER

 to continue the modulation correction.

The progress of the correction is monitored in the 

Status

 and 

Progress

 fields. Once the correction is

terminated, the 

Test

 status indication changes to 

HLT

 and the CMU is ready to perform an improved

wideband modulation analysis.

Important Note:

A warm-up time of at least 30 minutes is necessary to reach the normal operating temperature
of the CMU; otherwise the correction values might be impaired.

Checking the Corrected Measurement

The accuracy of the corrected modulation measurement can be checked by using a standard QPSK or
WCDMA signal from a signal generator. We propose the following equipment and test setup:

Equipment

Properties

Appropriate R&S device

Stock no.

Use

SMIQ03

1084.8004.03

QPSK

Signal generator

300 kHz to 3.3 GHz
QPSK / WCDMA

SMIQ03B
with option SMIQ-B47

1125.5555.03

WCDMA

Coax cable

N to N, approx. 1 m length

RF connection cable

0100.7670.10

RF connection
SMIQ –> CMU

Coax cable

BNC to BNC (50 

),

approx. 1.8 m length

RF connection cable

0100.6945.20

Synchronize CMU to 10 MHz
reference signal from SMIQ

Содержание CMU-Z6

Страница 1: ...1150 0199 12 03 1 Test and Measurement Division Operating Instructions Improvement of Wideband Modulation Analysis CMU Z6 1150 0001 02 Printed in the Federal Republic of Germany...

Страница 2: ......

Страница 3: ...ory environments or in the field and must not be used in any way that may cause personal injury or property damage You are responsible if the product is used for an intention other than its designated...

Страница 4: ...r unassociated documentation can result in misinterpretations and thus contribute to personal injury or material damage Basic safety instructions 1 The product may be operated only under the operating...

Страница 5: ...ccordingly 9 In the case of products of safety class I with movable power cord and connector operation is permitted only on sockets with earthing contact and protective earth connection 10 Intentional...

Страница 6: ...roperly replaced this can cause an explosion warning lithium cells Replace the battery or storage battery only with the matching Rohde Schwarz type see spare parts list Batteries and storage batteries...

Страница 7: ...a persona cosa pueda ser da ada El uso del producto fuera de sus fines definidos o despreciando las informaciones de seguridad del fabricante queda en la responsabilidad del usuario El fabricante no s...

Страница 8: ...definici n Por eso se debera tener en cuenta que las palabras de se al aqu descritas sean utilizadas siempre solamente en combinaci n con la correspondiente documentaci n y solamente en combinaci n co...

Страница 9: ...da En vista a la protecci n de la vida en desarrollo deber an ser protegidas personas embarazadas debidamente Tambi n las personas con un bypass pueden correr peligro a causa de la radiaci n electroma...

Страница 10: ...y la conexi n deber n ser efecutadas por un electricista especializado 20 En caso de que los productos que son instalados fijamente en un lugar sean sin protector implementado autointerruptor o simila...

Страница 11: ...ejemplo paredes y estantes 31 Las asas instaladas en los productos sirven solamente de ayuda para el manejo que solamente est previsto para personas Por eso no est permitido utilizar las asas para la...

Страница 12: ......

Страница 13: ...mode REF IN RF OUT DC Supply Rohde Schwarz CMU Z6 1150 0001 02 0 1 A 1 0 B 1 1 C MODE SETTING ON Input for 110 8 MHz reference frequency DC power supply Output for test signal LEDs indicating the ope...

Страница 14: ...with the 5 V DC power from the AUX connector at the rear panel of the CMU Use the two N to BNC cables to provide CMU Z6 with a 110 8 MHz reference frequency from RF3 OUT and to feed the generated test...

Страница 15: ...ormal operating temperature of the CMU otherwise the correction values might be impaired Checking the Corrected Measurement The accuracy of the corrected modulation measurement can be checked by using...

Страница 16: ...on measurement Perform a Preset of the SMIQ settings and a Reset of the CMU Set the SMIQ to comply with the CMU default analyzer settings FREQUENCY 1922 4 MHz LEVEL 0 dBm DIGITAL MOD STATE ON SELECT S...

Отзывы: