![Rohde & Schwarz 4091.7009.02 Скачать руководство пользователя страница 27](http://html1.mh-extra.com/html/rohde-and-schwarz/4091-7009-02/4091-7009-02_getting-started_1477973027.webp)
Operation
R&S
®
EB510
27
Getting Started 4091.7338.02 ─ 01
a) The BW +/- key changes the demodulation bandwidth and the dark blue
region displays the bandwidth in relation to the IF span.
b) The AFC key toggles automatic frequency control. When a peak is within the
demodulation bandwidth (3), AFC helps to track and keep the demodulation
frequency tuned to the frequency of the peak. AFC does not function in SSB
(USB or LSB) demodulation mode (4d).
c) The DET key changes the level detection mode (Average, Peak, Fast, RMS)
of the RF amplitude envelope.
d) The DEMOD +/- key changes the demodulation mode (CW, LSB, USB, ISB,
AM, FM, Pulse, PM, IQ). CW, USB and LSB demodulation modes are only
available with demodulation bandwidths (4a) of 9
kHz or narrower.
5. The SELECT Rollkey & Control (
chapter 2.1, "Front-Panel Tour"
allows changes to various receiver functions.
a) The MARK key enables the marker function. The "Marker" softkey bar allows
two sets of X markers and one set of Y markers to be set.
b) When the marker set is selected, turning the SELECT rollkey will move the
selected marker highlighted in green as illustrated above. The selected
marker can be toggled by pressing the "X1 / X2", "X3 / X4" or "Y1 / Y2"
softkey.
Graphical User Interface (GUI)