ZA15 External Attenuator
R&S
®
RT
‑
ZD10/20/30
31
User Manual 1410.4550.02 ─ 01
R&S RTO settings
●
Horizontal scale = 1 ms/div
●
Bandwidth = 20 MHz
●
Vertical scale = 100 mV/div
●
Trigger source = Ch1, Trigger mode = Auto, Run cont
●
ACQUISITION key > Decimation Mode = High res, Waveform arith-
metics = Average, Average Count = 10
●
"Meas" menu > Setup > Source = Ch1, Main Measurement = Mean,
State = On
●
"Vertical" menu > "Probe Setup" > "Ch" tab > "Attenuator RT-ZA15"
●
"Vertical" menu > Channels > Ch1 > Offset = 0 V, Position = 0 div
Adjustment
●
Set the voltage of the DC voltage source to 0 V.
●
Check the displayed value in the "Measurement Results" box: Mean =
V1.
●
Set the voltage of the DC voltage source to 30 V.
●
Check the displayed value in the "Measurement Results" box: Mean =
V2.
●
Using the small skrewdriver, turn the trimmer for DC common mode
rejection on the external attenuator until V2
=
V1.
Adjusting