Putting into Operation
R&S
®
RT
‑
ZD10/20/30
21
User Manual 1410.4550.02 ─ 01
As soon as the probe is connected to the oscilloscope and the settings are adjus-
ted, the waveform is shown for the channel to which the probe is connected.
To display probe-specific settings and information:
► On the "Vertical" menu, select "Probe Setup" and select the correct channel
tab.
2.1.3
AutoZero
This procedure performs an automatic correction of the zero error displayed on
the oscilloscope's screen. The waveform is set to 0 V on the horizontal centerline
of the oscilloscope.
To perform the AutoZero function using the micro button (see also:
1. Connect the probe to the R&S
RTO oscilloscope.
2. Configure the micro button function on the R&S
RTO:
"Vertical" menu > "Probe Setup" > "Ch" tab > "Micro button action" = 'Auto-
Zero'.
3. Short the signal pins and the ground pin together.
4. Press the micro button.
Alternatively, AutoZero can be started in the dialog box of the R&S
RTO:
► "Vertical" menu > "Probe Setup" > "Ch" tab > "Detect AutoZero"
Installation