17.7.2.2
Measurements results
You can query the statistical results using the
MEAS:STAT
commands.
To export statistical results to a csv file, use the
EXP:MEAS:STAT
commands. Note
that export of statistics is possible only remotely, but not in manual operation.
.................................................................................. 521
.................................................................................. 521
MEASurement<m>:RESult[:ACTual]?
...............................................................................522
.................................................................................... 522
...............................................................................522
.................................................................................523
................................................................................. 523
MEASurement<m>:RESult:WFMCount?
...........................................................................523
MEASurement<m>:STATistics:WEIGht?
...........................................................................523
MEASurement<m>:STATistics:VALue:ALL?
...................................................................... 524
MEASurement<m>:STATistics:VALue<n>?
....................................................................... 524
EXPort:MEASurement<m>:STATistics:NAME
....................................................................524
EXPort:MEASurement<m>:STATistics:SAVE
.................................................................... 525
EXPort:MEASurement:STATistics:ALL:NAME
................................................................... 525
EXPort:MEASurement:STATistics:ALL:SAVE
.................................................................... 525
MEASurement<m>:TIMeout[TIME]
<TimeoutTime>
Sets a waiting time, how long the instrument waits until it returns the measurement
results on
MEASurement:RESult?
.
After changing measurement or channel parameters, the instrument needs a setup
time of about 200
ms and the time of a valid acquisition (at least
12
*
horizon-
tal
scale
+
trigger period
). Only after this time, valid measurement results can be
returned.
You can define this waiting time manually, or let it set automatically with
Suffix:
<m>
.
The suffix is irrelevant.
Parameters:
<TimeoutTime>
*RST:
0
Default unit: s
MEASurement<m>:TIMeout:AUTO
<AutomaticMode>
The waiting time is set automatically, based on the horizontal scale. For details, see
Suffix:
<m>
.
The suffix is irrelevant.
Measurements