Remote control commands
R&S
®
RTM3000
515
User Manual 1335.9090.02 ─ 09
MEASurement<m>:AON
Starts the quick measurement.
Usage:
Event
MEASurement<m>:AOFF
Stops the quick measurement.
Usage:
Event
MEASurement<m>:ARESult?
Returns the results of the quick measurement.
Return values:
<Data>
List of values
Quick measurement results are listed in the following order:
PEAK (Vpp), UPE (Vp+), LPE (Vp-), CYCR (RMS-Cyc), CYCM
(MeanCyc), PER (T), FREQ (f), RTIM (tr), FTIM (tf).
Usage:
Query only
MEASurement<m>:ALL[:STATe]
Starts or stops the quick measurement and sets the status bit.
Suffix:
<m>
.
1..8
The suffix is irrelevant.
Parameters:
<State>
ON | OFF
*RST:
OFF
17.7.2
Automatic measurements
...........................................................................................515
...........................................................................................521
................................................................................................ 526
17.7.2.1
Measurement settings
...........................................................................................516
................................................................................................516
........................................................................................... 518
...................................................................................518