Command Reference
R&S
®
ZNA
1051
User Manual 1178.6462.02 ─ 12
Example:
See
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:
Manual operation:
See
8.3.14.3
[SENSe<Ch>:]CDLL...
Adds, removes and configures
These commands allow you to add/remove external DLLs to/from the firmware. It gives
access to their configuration, tasks, and traces.
[SENSe:]CDLL:ADD
<DllPath>
Tells the analyzer firmware to load a custom (external) DLL.
The DLL and its accompanying files must be packed in a zip archive. It will be loaded
on every subsequent firmware start, until it is removed using
.
Note that adding/removing custom DLLs is only possible if the firmware is run as
administrator.
Setting parameters:
<DllPath>
Full path to the zip archive including the DLL (at the top-tevel).
Usage:
Setting only
Manual operation:
See
[SENSe:]CDLL:LIST?
Queries the loaded custom (external) DLL files.
The return value is a single string, comprising a comma-separated list of DLL names,
without the
dll
extension.
Use
to load/unload a custom DLL..
Usage:
Query only
Manual operation:
See
[SENSe:]CDLL:REMove
<DllName>
Tells the analyzer firmware to remove a custom (external) DLL that was previously
added using
.
Note that adding/removing custom DLLs is only possible if the firmware is run as
administrator.
Setting parameters:
<DllName>
Name of a loaded DLL (see
Lowercase string, without
dll
extension
SCPI Command Reference