Preparing for Use
R&S
®
ZVA-Z90E / -Z110E
11
Getting Started 1307.7197.62 ─ 08
Risk of instrument damage
An unsuitable operating site or device condition can damage the converter. Before
switching it on, observe the information on appropriate operating conditions provided in
the data sheet. In particular, ensure the following:
●
Converter covers are in place and all fasteners are tightened.
●
Ventilation openings are unobstructed.
●
The converter is dry and shows no condensation.
2.4.1
Unpacking the Unit and Checking the Shipment
When your receive the converter, please take the following steps:
1. Unpack the converter and the other contents of the wooden shipping box.
2. Check the equipment for completeness using the delivery note and the accessory
lists for the various items.
3. Remove the protective cap from the test port adapter at the front of the instrument.
4. Carefully inspect the frequency converter for any damages during shipment.
If you notice any damage, immediately contact the carrier who delivered the instru-
ment.
Risk of damaging waveguide flanges
The waveguide flanges of the converter and of the test port adapters must be protec-
ted against mechanical damage. Furthermore the waveguides must be shielded from
dust.
Protect the waveguide flange of the converter by leaving a test port adapter mounted.
When the converter is not in use, attach one of the included protective caps to the
adapter. Also attach protective caps to the second test port adapter.
Retain the wooden box and use it to store the converter and its accessories while not
in use. In particular, use the box if the converter needs to be transported or shipped,
e.g. if returned or sent in for repair.
Putting the Converter into Operation