R&S FSH
Two-port measurements with the tracking generator
1145.5973.12 4.101
E-11
Settings for detection of the R&S FSH-Z2 and R&S FSH-Z3
Operation:
Press the SETUP key.
Select the HARDWARE SETUP softkey.
Select ACCESSORY from the menu by using the
rotary knob or the cursor keys and confirm the entry
with the ENTER key.
The R&S FSH opens a menu where different modes
for detecting the R&S FSH-Z2 or R&S FSH-Z3 VSWR
bridge can be selected.
Select AUTO DETECT or BRIDGE FSH-Z3 or
BRIDGE FSH-Z2 from the menu by using the rotary
knob or the cursor keys and confirm the entry with
the HARDWARE SETUP softkey or the ENTER key.
If you select AUTO DETECT, the VSWR bridge is
detected automatically as soon as the control cable
has been connected to the probe power socket on the
R&S FSH.
Note:
If you use the R&S FSH-Z2 bridge, removal of this
VSWR bridge will not always be detected
automatically. In this case, briefly change to another
measurement function or the menu item NONE in
order to update the instrument status.
If you select BRIDGE FSH-Z2/-Z3, the VSWR bridge is
detected by default. This setting may be useful if you
only work with the R&S FSH-Z2/-Z3 connected and do
not want to waste time on automatic detection.