CRAFT™ Oocyte Aspiration Pumps – Service Manual

 

25

 

Copyright© 2012-19 

Rocket Medical plc. . 

All rights reserved. 

ZDOCK226 

Rev:9 

25/02/19 

 

 

 

 
 

 

WARNING: 
ELECTRIC SHOCK 
HAZARD.

 

Do not immerse the 
device.

 

 

 

 

WARNING:  

Do NOT

 

attempt to sterilise the 
device

 

 

 

 

WARNING: 

Protect the 

device from ingress of 
liquid. Should any liquid 
enter the device, 
discontinue use 
immediately and refer 
to an authorised service 
agent

 

 
 
 
 
 

 

 

WARNING: 

No user 

serviceable parts 
inside.

 

 

 

 

HAZARD:

 

Do NOT include used 
consumables as these 
pose a significant 
contamination risk

 

 
 

 

IMPORTANT

 

A decontamination 
certificate MUST be 
included with every 
returned pump.

 

Repair or servicing 
cannot be commenced 
unless the service 
agent is in possession 
of this certificate

 

10.

 

CLEANING THE PUMP CASING

 

At the end of each clinical session, turn off the device at the front panel 

 

and disconnect 

the device from mains power supply 

Using an aqueous solution of 70% alcohol (IMS or isopropyl BP), moisten a cloth and wipe 
all  external  surfaces  of  the  device.  If  the  surface  has  become  contaminated  with 
proteinaceous material, remove with a light detergent solution before surfacing cleaning with 
alcohol. 

Do NOT use a 100% alcohol or any other solvent to clean the device as this may cause 
damage to the casing surface and display. 

Prevent any fluid from entering the device. 

 
 
 
 
 
 

11.

 

YEAR OF MANUFACTURE:

 

 

 
 

12.

 

RETURNING THE PUMP FOR FACTORY 

REPAIR OR SERVICE:

 

All devices to be returned must be prepared as described below for the protection of the 
servicing team and for safety during transport. 

1.  Surface clean the pump as described in the Section above 

2.  Seal in a plastic bag and seal within a second plastic bag. 

3.  Place in the original packaging. 

4.  Enclose the following information: 

• 

Contact name 

• 

Centre address 

• 

Decontamination Certificate

 

• 

Description of the fault or service required 

• 

Accompanying Order to authorise servicing 

– contact your local Customer 

Services Team for details 

Units  manufactured  before  2014: 

The  year  in  which  the  device  was 

manufactured is  indicated  by  the first 2 numbers of the serial number. 
For  example:  a  serial  number  starting 

11

180776  indicates  the  device 

was manufactured in 2011 

For units manufactured after 01/04/14, the year of manufacture is shown 
on the rear rating plate label opposite the model number. 

Содержание CRAFT

Страница 1: ...SERVICE MANUAL Rocket CRAFT and DUOVAC Oocyte AspirationPumps ...

Страница 2: ...ON CRAFT SUCTION DUO VAC PUMPS 11 4 11 SYMBOLS USED ON R57685 PATIENT CONNECTION SET 11 5 GENERAL INFORMATION 12 5 1 General Description 12 5 2 Indications 12 5 3 Contraindications 12 5 4 References 12 6 ANNUAL SERVICE 13 6 1 Replacing the filter element 13 6 2 Replacing the pump valve diaphragm set 15 6 3 Reassembling the valve set 18 7 REPAIRS REPLACEMENTS 20 7 1 Replacing the DUO VAC Footswitch...

Страница 3: ...ontrol Dial clockwise to increase anticlockwise to decrease the set value 3 Vacuum Display mmHg 1 4 Footswitch connection ports 5 Water trap connection port recommended for use with R57685 Rocket Craft Pump Water Trap Sets 6 Water Trap Set comprising glass bottle bung angled connectors silicone tube set Recommended for use with R57685 Rocket Craft Pump Water Trap Sets 7 Silicone tube set with meta...

Страница 4: ...6 Rev 9 25 02 19 2 2 Rocket CRAFT DUO VAC Suction Pump Internal View 1 Vacuum gauge 0 700mmHg normal range 50 440mmHg 1 2 Illuminated O I Mains Power On Off 3 Motor 220VAC 50Hz 110VAC 60Hz model dependant 4 Vacuum outlet port 5 Pump head including valve housing 6 Internal silencer with filter chamber 7 Main power input ...

Страница 5: ... Vacuum 440mmHg 1 control button 4 Vacuum Control clockwise to increase anticlockwise to decrease the set value 5 Water trap connection port recommended for use with R57685 Rocket Craft Pump Water Trap Sets 6 Disposable Water Trap Set comprising glass bottle bung angled connectors PVC tube set 7 Footswitch connection port 8 Medium Vacuum Standard 50 250mmHg 1 footswitch Footswitch supplied on Rock...

Страница 6: ...CK226 Rev 9 25 02 19 2 4 Rocket CRAFT Suction Pump Internal View 1 Vacuum gauge 0 700mmHg normal range 50 440mmHg 1 2 Illuminated O I Mains Power On Off 3 Motor 220VAC 50Hz 110VAC 60Hz model dependant 4 Vacuum outlet port 5 Pump head including valve housing 6 Internal silencer with filter chamber 7 Main power input ...

Страница 7: ...st tube for use with oocyte needle sets is B D Falcon test tube No 2001F 17 x 100mm 3 1 INDICATIONS For the generation of high vacuum low volume suction between 10 400 mmHg 1 to permit the aspiration of follicular fluid oocytes and ovarian fluid as part of the treatment of infertility relating to IVF and other gynaecological procedures 3 2 CONTRAINDICATIONS Not intended for use where ovarian aspir...

Страница 8: ...MATION 4 1 COPYRIGHT This manual contains information that is subject to copyright All rights reserved This manual should not be photocopied duplicated or distributed completely or in part without the written approval of Rocket Medical plc 4 2 MODEL NUMBERS Rocket Craft Suction Pump Complete 110v R29654 Rocket Craft Suction Pump Complete 240v R29655 Rocket Craft Duo Vac Suction Pump 240v R29660 Ro...

Страница 9: ...invalidate the Warranty UK European Service Agents Hunter Scientific Limited Unit 1 Priors Hall Widdington Saffron Walden Essex CB11 3SB Tel 44 0 1799 541 688 Fax 44 0 1799 541 703 E service hunterscientific com W www hunterscientfic com UK Customer Services Rocket Medical plc Sedling Road WASHINGTON NE38 9BZ ENGLAND Tel 44 0 191 419 6988 Fax 44 0 191 419 6989 Email customerservices rocketmedical ...

Страница 10: ... degree of protection against harmful interference found in typical medical installations Medical electrical equipment requires special precautions regarding EMC and the device must be installed positioned and operated according to the instructions contained in this manual to ensure continued electromagnetic compatibility The device must be operated according to the instructions contained in this ...

Страница 11: ...FT SUCTION DUO VAC PUMPS These tables describe their respective meanings Mains Power ON OFF This device is Type B CE Mark 4 11 SYMBOLS USED ON R57685 PATIENT CONNECTION SET Read the Instruction for Use before connection and use Device is for Single Use Only Device is sterilised by Ethylene Oxide This device is not manufactured with natural rubber latex CE Mark ...

Страница 12: ...table collection test tube for use with oocyte needle sets is B D Falcon test tube No 2001F 17 x 100mm 5 2 Indications For the generation of low flow high vacuum between 10 440 mmHg 1 to permit the aspiration of follicular fluid oocytes and ovarian fluid as part of the treatment of infertility relating to IVF and other gynaecological procedures 5 3 Contraindications Not intended for use where ovar...

Страница 13: ...e filter element 1 Lay the pump on a clean surface protecting the casing from scratches 2 Remove 4 screws from the casing underside and retain 3 Carefully separate the case halves ensuring the front and rear fascia plates remain located in the lower shell 4 Locate filter housing 5 Using fingers lift the top away from filter chamber CAUTION Avoid the use of sharp tools as these may damage the O rin...

Страница 14: ...al plc All rights reserved ZDOCK226 Rev 8 25 02 19 6 Life out the filter assembly and remove the filter 7 Fit new replacement filter P N E05 000 8 Replace filter ensuring the top seats securely into the chamber 9 Avoid trapping or pinching the O ring 10 Check the top is correctly seated ...

Страница 15: ... Service Manual 6 2 Replacing the pump valve diaphragm set 1 Locate the pump silencer housing and carefully remove with a gentle twisting movement 2 Using a diagonal method remove the 4 screws securing the pump valve head 3 Lift the valve head away from the pump body to reveal the pump head gasket and diaphragm assembly ...

Страница 16: ...TANT 4 Undo and remove the centre Pozi drive screw to release the valve plate 5 Carefully remove valve plate from head moulding with a twisting motion to separate the two halves 6 Valve gasket and valve plate Note the orientation of the cut outs on the valve gasket in relation to the valve plate 7 Lift off pump head gasket and retain safely ...

Страница 17: ...mps Service Manual IMPORTANT 8 Undo and remove the centre Pozi drive screw from the pump diaphragm retaining plate 9 Retain retaining plate and screw discard old diaphragm 10 Replace diaphragm with P N DO1 053 11 Re assemble and tighten sufficient to secure the retaining plate DO NOT OVERTIGHTEN or use excessive force ...

Страница 18: ...ANT 6 3 Reassembling the valve set 1 Ensure the valve plates are orientated as in the image right IF THIS ORIENTATION IS REVERSED THE VALVE AND PUMP WILL FAIL TO OPERATE 2 Replace the valve gasket with P N VO1 002 Note the orientation of the cut outs on the valve gasket in relation to the valve plate 3 Reassemble the valve head and tighten the centre screw ...

Страница 19: ...MPORTANT 4 Reseat the pump head gasket P N GO1 001 5 Replace the valve head assembly onto the gasket ensure the holes are correctly aligned and the gasket with seated correctly 6 Tighten the valve head screws in a diagonal pattern to ensure the pump head is correctly seated without distortion 7 Re fit the pump silencer to the head module ...

Страница 20: ...e subject to damage through misuse or repeated over stressing of the retaining clips on connection of the footswitch Replacement is a straight forward operation 1 Remove the upper casing See Section 6 2 Lift the front panel out of the retaining grooves on the lower case 3 Remove tubing by carefully pulling from connector 4 Loosen and remove securing nut 5 Replace connector locate and tighten secur...

Страница 21: ...s on the lower case 3 Loosen and remove the securing nuts or screws ringed right 4 Draw the gauge through the front panel whilst supporting the mounting plate 5 Detach the gauge from the centre tube by carefully pulling the tubing from the rear connection 6 Replace the gauge reconnect the centre tubing and insert through the front panel 7 Replace gauge mounting plate and tighten securing nuts or s...

Страница 22: ...ocket CRAFT Oocyte Aspiration Pumps Service Manual 22 Copyright 2012 19 Rocket Medical plc All rights reserved ZDOCK226 Rev 9 25 02 19 IMPORTANT 7 3 Bottle Bracket Replacement using Service Kit B02 190S ...

Страница 23: ...e tube to the test gauge 7 Activate the Standard Vacuum footswitch observe the green light is illuminated 8 Rotate the control knob clockwise to increase to the desired value on the gauge See table below 9 To decrease the value rotate anticlockwise 10 When the desired value is reached release the footswitch and tubing set occlusion WARNING Ensure the vacuum levels are within the specified accuracy...

Страница 24: ...TRIC SHOCK HAZARD The equipment is to be used only with electrical systems complying with all IEC CEC and NEC requirements IMPORTANT 1 Two 0 5A fuses are located in the main power socket mounted on the rear panel 2 Using a small screwdriver carefully ease out the fuse carrier to reveal the fuses 3 To ensure continued protection REPLACE WITH FUSES OF IDENTICAL RATING ...

Страница 25: ...h proteinaceous material remove with a light detergent solution before surfacing cleaning with alcohol Do NOT use a 100 alcohol or any other solvent to clean the device as this may cause damage to the casing surface and display Prevent any fluid from entering the device 11 YEAR OF MANUFACTURE 12 RETURNING THE PUMP FOR FACTORY REPAIR OR SERVICE All devices to be returned must be prepared as describ...

Страница 26: ...y and refer to an authorised service agent 14 TRANSPORTATION The device must be transported in temperatures 5 C to 35 C The device is FRAGILE and must be transported in its original packaging to ensure protection If the original packaging is not available please contact your local Customer Services Agent who will provide replacement packaging Dimensions W 264mm H 124mm D 164mm Weight Unit 1 9Kg Fo...

Страница 27: ... Medical plc or other than in accordance with written instructions provided by Rocket Medical plc or altered by anyone other than Rocket Medical plc or if the Products have been subject to misuse negligence or accident Rocket Medical plc s sole and exclusive obligation and Buyer s sole and exclusive remedy under the above warranties is limited to repairing or replacing free of charge at Rocket Med...

Страница 28: ...n against electric shock Type B 16 2 SPECIFICATIONS Power Input to Pump 220 240 VAC 50Hz 0 12A Maximum current 2 5A Consumption 0 6KW h Environmental conditions Temperature 5 C to 35 C Atmospheric Pressure Range 700 1060hPA Dimensions W 264mm H 124mm D 164mm Weight Unit 1 9Kg Foot Switch 0 51Kg Vacuum Ranges Medium vacuum 20mmHg to 200mmHg in 20mmHg increments High vacuum 400mmHg Vacuum Range Accu...

Страница 29: ...Rocket CRAFT Oocyte Aspiration Pumps Service Manual 29 Copyright 2012 19 Rocket Medical plc All rights reserved ZDOCK226 Rev 9 25 02 19 17 DRAWINGS 17 1 GENERAL ASSEMBLY ...

Страница 30: ...Rocket CRAFT Oocyte Aspiration Pumps Service Manual 30 Copyright 2012 19 Rocket Medical plc All rights reserved ZDOCK226 Rev 9 25 02 19 17 2 GENERAL ELECTRICAL ...

Страница 31: ...Rocket CRAFT Oocyte Aspiration Pumps Service Manual 31 Copyright 2012 19 Rocket Medical plc All rights reserved ZDOCK226 Rev 8 25 02 19 17 3 GENERAL INSTALLATION ...

Отзывы: