CRAFT™ Oocyte Aspiration Pumps – Service Manual
20
Copyright© 2012-19
Rocket Medical plc. .
All rights reserved.
ZDOCK226
Rev:8
25/02/19
7.
REPAIRS & REPLACEMENTS
7.1
Replacing the DUO-VAC Footswitch Connectors
The front panel connectors of
the DUO-VAC can be subject
to damage through misuse or
repeated over-stressing of the
retaining clips on connection
of the footswitch.
Replacement is a straight
forward operation
1. Remove
the
upper
casing. (See Section 6)
2. Lift the front panel out of
the retaining grooves on
the lower case
3. Remove tubing by
carefully pulling from
connector.
4. Loosen and remove
securing nut.
5. Replace connector,
locate and tighten
securing nut.
6. Re-fit tubing by pushing
over connector inlet.