Rocket
CRAFT™ Oocyte Aspiration Pumps – Service Manual
12
Copyright© 2012-19
Rocket Medical plc. .
All rights reserved.
ZDOCK226
Rev:8
25/02/19
5.
GENERAL INFORMATION:
5.1. General Description:
The Rocket
CRAFT™ Oocyte Aspiration Pumps have been developed to provide smooth,
low volume/high vacuum at a pre-determined negative pressure. Vacuum is activated by a
foot operated toggle air switch controlled by the surgeon performing the oocyte collection.
The range of vacuum is infinitely variable from 10-200mmHg
-1
in medium vacuum mode and
at a pre-set 440mmHg
-1
in high vacuum mode.
Rocket
CRAFT™ Oocyte Aspiration Pumps require either a water trap set (supplied) or
R57685 Disposable Rocket
CRAFT™ Water Trap Sets, supplied separately, sterile and for
single patient use.
A suitable collection test tube for use with oocyte needle sets is B.D. Falcon test tube No.
2001F, 17 x 100mm.
5.2. Indications:
For the generation of low flow/high vacuum between 10-440 mmHg
-1
to permit the aspiration
of follicular fluid, oocytes and ovarian fluid as part of the treatment of infertility relating to IVF
and other gynaecological procedures.
5.3. Contraindications:
Not intended for use where ovarian aspiration or the aspiration of ovarian fluid is
contraindicated. For short term operation only. This device is not indicated for continuous
drainage.
5.4. References:
Craft I, McLeod F. Edmonds K, ‘Human embryo transfer technique’. Lancet 1961 ii 1104-5
Craft I, Diahanbakch O. McLeod F et al ‘Human pregnancy following oocyte and sperm
transfer to the uterus.’ Lancet 1992 i 1031-3
Craft I, (1984) ‘Clinical Methodology’ British Journal of Hospital Medicine 90-102
Reeves G, Scott R T, et al (1989) Journal of Assisted Reproduction and Genetics Volume
6, Number 6 / December, 1989