![Roche OMNI C Скачать руководство пользователя страница 126](http://html1.mh-extra.com/html/roche/omni-c/omni-c_instructions-for-use-manual_1472343126.webp)
7 Trouble shooting
7-6
Instructions for Use, Roche OMNI C, Rev. 4.0, December 2002
7.2.9 3012 User system stop
Cause
•
The automatic fluidic procedure completion of some system stops can be interrupted by
a User system stop (by pressing the "Stop" button), e.g. in order to get immediate access
to the "More functions" button.
Remedy
•
Press the "OK" button (terminate the "User system stop")
•
In case of recurrence, call customer service!
Consequence actions
•
Aspirate Mix1
7.2.10 3013 Fluid pack switch
Cause
•
The docking mechanism of Fluid Pack C3 has been opened (micro switch activated)
•
The Fluid Pack C3 docking mechanism micro switch is defective
Remedy
•
Close docking mechanism Fluid Pack C3
•
Check micro switch ("More functions > System > Test > Control sensors > Monitoring
sensors"),.
•
In case of recurrence, call customer service (change if necessary)!
Consequence actions
•
Auto-preparing of Fluid Pack C3 solutions
7.2.11 3014 Fill level alarm
Cause
•
The solutions C1, C2 and/or C3 are empty (below alarm level) or are set to "empty"
•
The Waste container W is full (above alarm level)
•
The expiry date of the solutions is exceeded
•
The on-board lifetime of the solutions is exceeded (C1/C2 = 28 days; C3 = 42 days)
Remedy
•
Change solutions C1, C2 and Pack C3
•
Change or empty the Waste container W according to the instructions
•
In case of recurrence, call customer service!
Consequence actions
•
C1 changed: prepare solution C1, fill FMS reservoir, conductivity calibration
•
C2 changed: prepare solution C2, fill FMS reservoir, conductivity calibration
•
C3 changed resp. C3 docking mechanism micro switch operated: prepare C3 solutions
•
Waste container changed resp. micro switch operated: Waste container fill level
measurement
Содержание OMNI C
Страница 6: ......
Страница 44: ...1 Introduction 1 36 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Страница 46: ...2 Specifications 2 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Страница 56: ...2 Specifications 2 10 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Страница 58: ...3 Calibration 3 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Страница 62: ...3 Calibration 3 4 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Страница 64: ...4 Measurement 4 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Страница 76: ...4 Measurement 4 12 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Страница 78: ...5 Quality control 5 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Страница 118: ...6 Maintenance 6 24 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Страница 120: ...7 Trouble shooting 7 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Страница 132: ...7 Trouble shooting 7 12 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Страница 134: ...8 Operating modes 8 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Страница 148: ...8 Operating modes 8 14 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Страница 152: ...Index 4 Instructions for Use Roche OMNI C Rev 4 0 December 2002...