Manual input of the parameters

  

Edit Setup

 opens the setup screen. When a setup screen is 

opened, the input marker is located initially on the 

Dilution

 

input fi eld.

  

The current input position is indicated by a fl ashing, horizontal 
line below the characters which can currently be changed. 
From this point, you can use the 

MENU

 key to move forwards 

or backwards between the input fi elds. 

  

To move forwards, press the 

MENU

 key briefl y. The input 

maker skips forward to the next available input position.

 

 To move backwards, keep the 

MENU

 key pressed until the 

input marker reaches the position you require. Both methods 
move you to the start or the end of the screen when the last 
or the fi rst parameter respectively is reached.

   

When inputting numbers, you can move forwards from 
number to number using the 

NEXT

 key. 

 

 To increase or decrease a number selected using 

NEXT

use the 

 and 

 keys. For all other parameters, the 

 and 

 keys can be used to select various pre-defi ned values. 

This also applies for the individual characters of the setup 
name. 

  

You must save your entries if you want the settings to be 
available next time you switch on CASY Model DT. To do this, 
press 

ENTER

 and you will be forwarded to the Saving Setup 

where the existing setups are listed. Use 

NEXT

 to select one 

of the 20 setups and then press 

ENTER

. The setup currently 

loaded will be offered as a default setting. You may select set-
ups that have already been saved and therefore tagged with * 
and also unused setups. If you select a setup tagged with *, its 
stored settings will be overwritten with your new settings. If you 
select an unused setup, it will appear tagged with * and will 
be available for selection next time 

Select Setup

 is opened. 

SETUP NUMBER: 00 

 

NAME:

 

Default

Capillary: 150 

μ

m

Sample Volume:

1 x 200 

μ

l

Dilution:

1.000e+00

Left Cursor:

3.3 

μ

m

Right Cursor:

50.0 

μ

m

Display Mode:

Standard

Total Counts/ml:

Off

Interface:

Parallel

Print-Mode:

Manual/PF On

USE MENU AND NEXT TO STEP. USE +/- TO EDIT. USE 
CANCEL TO LEAVE SCREEN OR ENTER TO CONFIRM CHANGES

    

 If you do not wish to save your settings, press 

CANCEL

. The 

new settings are used until you select an alternative setup 
or switch off your CASY Model DT. They will then no longer 
be available. As a setup name for unsaved settings, 

NOT 

SAVED

 appears and the setup number is replaced by two 

forward slashes //.

5. 

Monitoring the background.

1

Load or create a setup with a dilution factor of 1 and cursor positions covering the entire size range (at delivery Setup No 00) 

2

Place a CASY

 

cup containing CASY

 

ton below the measuring capillary.

3

Press and hold the 

CLEAN

 key down until the clean counter indicates 3. Releasing the key triggers three clean cycles.

4

Place a new CASY

 

cup containing CASY

 

ton below the measuring capillary.

5

Press the 

START

 key to initiate a measurement.

 For the used 150 µm measuring capillary the background 
should be < 100 Counts/ml (in the entire size range offered). 
If the background succeeds 100 Counts/ml, just measure 
another CASY

 

cup containing freshly fi ltered  CASY

 

ton 

without running CLEAN in between. In general the use of 
CASY

 

ton fi ltered by a sterile fi lter with a pore size 

 1,2 µm 

is recommended. 

 

   Nearly fault-free work is only ensured when a cleaning routine as described in the Operator Manual (5.1. Weekly cleaning 

routine) is performed at least once a week!