Measuring beads used by the 454 Sequencing System.

6. Setup 

selection

   

Load the appropriate setup in the Setup selection list 
(as described in 4.) or if not already available create and save 
a setup for the beads (as described in 4.) with the following 
parameters:

SETUP NUMBER: 01 

 

NAME:

 

Beads

Capillary: 150 

μ

m

Sample Volume:

3 x 400 

μ

l

Dilution:

3.333e+03

Left Cursor:

 15.00 

μ

m

Right Cursor:

25.00 

μ

m

Display Mode:

 Standard

Total Counts/ml:

Off

Interface:

Parallel

Print-Mode:

Manual/PF On

USE MENU AND NEXT TO STEP. USE +/- TO EDIT. USE 
CANCEL TO LEAVE SCREEN OR ENTER TO CONFIRM CHANGES

7. Preparing 

samples.

   Rotate the recess in the sample platform until it points 

forwards and remove the CASY

 

cup from below the 

measuring capillary.

   Place the sample to be measured below the measuring capillary 

and ensure that the external platinum electrode is immersed 
in the sample.

   Rotate the sample platform further to the side.

   Press 

START

 to perform a measurement using the current 

parameters. CASY Model DT automatically works through all 
the necessary steps. After a short time, the results for your 
sample appears in the display.

   Since the beads used by the 454 Sequencing System are in the 

size range between 15 and 25 µm the parameter 

Counts/ml 

15.0 – 25.0 µm:

 1,140e+07 indicates the measured bead 

concentration per ml in the bead suspension to be examined 
(used in 7.). 

9. 

Changing samples and fi nishing measurements

    

 Provided you measure samples of the same type in a similar 
concentration range, you do not necessarily have to fl ush 
the system when samples are changed. If you expect larger 
concentration differences or wish to perform measurements 
with different sample types, before a new measurement, place 
a CASY

 

cup containing freshly fi ltered CASY

 

ton below the 

measuring capillary and fl ush the system using the 

CLEAN

 

function.

    

 Before switching off the CASY Model DT monitor again the 
background (see 5.).

SET:

Beads-01

SN:

DT-2XX-9999

CAP:

150 

μ

m

DATE: 15.09.2009

SVOL: 3x 400 

μ

l

L.CURSOR:

 15.0 

μ

m

TIME: 10:15

DIL:

3.333E+03

R.CURSOR:

 25.0 

μ

m

Conc. Range:

OK

Counts:

4533

Counts/ml 

3.3-15.0 

μ

m:

 

1.127e+06

Counts/ml

 

15.0-25.0 

μ

m:

 

1.140e+07

Counts/ml 

25.0-100.0 

μ

m:

 

3.684e+05

For life science research only. Not for use in diagnostic procedures.
CASY and 454 are trademarks of Roche.

   Fill a CASY

 

cup with 10 ml CASY

 

ton (for highest accuracy 

ensure that exactly 10 ml CASY

 

ton are pipetted into the 

CASY

 

cup).

   Add 3 µl of the homogenized bead suspension to be examined 

(for highest accuracy ensure that exactly 3 µl are pipetted into 
the 10 ml CASY

 

ton).

   Secure the lid and mix the sample by tilting the CASY

 

cup three times. Carefully avoid the formation of bubbles or 
foam!

8.  

Measuring samples and displaying the measuring results

25381209 CASY Counter Model DT-Operating Instructions.indd   4

25381209 CASY Counter Model DT-Operating Instructions.indd   4