prietest TOUCH
USER MANUAL
Version No.: 2.622A
Effective Date: July, 2010
Page 34 of 52
9.3. KINETIC
Multiple readings are taken at set temperature, at regular intervals and change in absorbance per minute is
calculated. Concentration is calculated from the factor fed by the user or by using standard.
PROGRAMMING / ADDING a NEW TEST (Refer chapter 8)
Enter all the test parameters in Kinetic mode
KINETIC
Mode screen before programming would look like the above screen
Refer chapter11 for entering the test parameters .On completion, touch “SAVE” to save the programmed test
Recalling /Running (Operating) a pre-programmed test. (Refer chapter 8 )
For example:
.
Name
: Test Name,
Mod:
KIN
36.45
Pri.
: 340 ,
Sec. :
0
ESC
Temp
: 37C ,
KF
: 1.000
EDIT
Vol:
500ul
,
Unit:
mg/dl
RUN
Lag :
5
, Read:
10
QCR
Blank :
Y,
QC
: N,
Normal:
Y
STD:
N,
Factor:
1746
Limit Set:
To change values press EDIT
Name
: ,
Mod:
KIN
36.40
Pri. : 340
,
Sec. :
0
ESC
Temp
: 37C ,
KF
: 1.000
Vol:
300ul
,
Unit:
No –Unt
ADD
Lag :
0
, Read:
0
Blank:
N,
QC
: N,
Normal:
N
SAVE
STD:
N,
Factor:
0.000
Limit Set:
KINETIC
Mode screen (After recalling a saved test.)
To change values, touch”
EDIT
”
To run the programmed test, touch “
RUN
”
To print, touch
“PRINT”
(Printing is optional)
To escape, touch “
ESC
”
Test Name:
ESC
PID
RUN
Aspirate Reference
Press ASP Switch to sip
At this stage,
the analyser will wait till the temperature of
cuvette reaches the set temperature. The instrument
prompts for REFERENCE, and prompts the user to place
the reference solution (distilled water) below the tube and
to press the aspiration switch. The instrument aspirates
the solution and automatically adjusts the reference.
(Note :To abort temperature setting, touch “
ESC
” once)
(In cuvette mode when the instrument prompts reference
“close the lid and Press RUN” before inserting sample
cuvette)
Test Name:
ESC
PID
RUN
Wash
Run QC
Read Sample.
“Read sample” on display indicates the analyser is ready
for reading samples.
The user has to place the sample solution below the tube
and press the aspiration switch. The instrument
aspirates the sample and displays the readings/results.
To continue reading more samples do the same.
(Note :At this stage the operator can enter the Patient
Identification .To enter Patient ID ,touch”
PID
”. Entering
PID is optional. )
“Initial OD” in Limit Set is kept mandatory. Test
will not get saved if initial OD limit is not
entered. In such case it will give a message
“Enter Initial OD in limit Set”
The High/Low
option for Initial OD in Limit Set test parameter has
been removed.
Name
: KIN
Mod:
KIN
36.40
Pri. : 340
,
Sec. :
0
ESC
Temp
: 37C ,
KF
: 1.000
Vol:
300ul
,
Unit:
No –Unt
ADD
Lag :
0
, Read:
0
Blank:
N,
QC
: N,
Normal:
N
SAVE
STD:
N,
Factor:
0.000
Limit Set:
36.72
Init. OD: 0.000
ESC
Max Delta/Min: 0.000
Rgnt. Linearity: 0.000
Numeric value:
1 2 3 4 5 6 ENTER
7 8 9 0 . CLR