TECHNICAL FILE
2500W HMI FOLLOW SPOT
1. USING THE "AUTO" MODE .
message displayed :
RJ LAMP LIGHTING
TRY FAILURE
1.1 The ballast and the CAD are on ;
the "lamp on" command is given,
but the lamp does not switch on.
Check that the lamp is in good
condition, in any doubt : change it.
(counter lamplife : page 6/7 - manual CAD)
Check the mains voltage >200 V.
message displayed :
RJ LAMP LIGHTING
TRY FAILURE
Check that the lantern bonnet is well
closed.
Check that the cable is well connected.
(refer to Tech. File page 14 & 24 )
Check that the correct cable has
been used.
message displayed :
RJ LAMP LIGHTING
TRY FAILURE
No ignition sound.
Check that the ignition is well adjusted
( refer to Tech. File page 22 )
ignition sound.
Contact the RJ services.
The ballast does not work.
the mains checklight is not on
Check the mains.
Check that the circuit breaker
is engaged.
Check the CAD fuse.
the mains checklight is on
the CAD 900 is not on
the CAD still does not work :
Contact the RJ service.
Contact the RJ service.
Message displayed :
RJ LAMP ERR.
NON-FUNCTIONING DIAGNOSIS.
The CAD fuse blows out while in use.
Disconnect the shutter connection
and then change the fuse.
Switch on
the ballast.
If the fuse blows again - change
the CAD: ( refer to Tech. File p. 43 ).
The CAD stays on - check the
shutter cable connection.
(refer to Tech. File page 14 & 24 )
Reconnect the shutter cable :
If the problem comes back : Change
the whole block.
( refer to Tech. File page 34 )
Restart the
"LAMP ON"
procedure.
Restart the
"LAMP ON"
procedure.