background image

Table section 2 / maintenance measures, general  
 
 

 
 
PGC9300 Maintenance Logbook  ·  EN07  ·  2018, June 26th 

18 

Maintenance measure 

 

Performed 

by

 

 

Performed 

on

 

 

 

 

 

Measures performed 

 

Routine maintenance 

 

Table 3 

Software update 

 

 

Metrological check 

 

Table 4 

(Re)calibration 

 

Table 4 

Defect 

 

See below 

 
 

Description of problem 

Measure 

 

 

 

 

Description of problem 

Measure 

 

 

 
 

Signature 

 

 

Содержание PGC 9300

Страница 1: ...Process Gas Chromatograph PGC 9300 MAINTENANCE LOGBOOK Reliable Measurement of Gas Status 2018 JUNE 26TH Version 07...

Страница 2: ...ment may serve as reference for translations into other languages Please use in case of any uncertainties the German version as main reference Note Unfortunately paper is not updated automatically whe...

Страница 3: ...PGC year of manufacture Meter serial number GC 9300 manufacturer number GC 9300 year of manufacture GC 9300 software version Approved for Calorific value Gas quality Operator Station Commissioning Typ...

Страница 4: ...1 1 1 2 Cylinder pressures 2 1 2 Maintenance by service technicians 2 2 Manufacturer s requirements 3 2 1 Annual maintenance of the process gas chromatograph 3 2 2 Metrological check re calibration 3...

Страница 5: ...ion 1 of this maintenance logbook must be documented during every inspection Routine maintenance should be performed once a year by qualified personnel according to the manufacturer s instructions and...

Страница 6: ...ould be checked for Carrier gas 1 Carrier gas 2 if applicable Internal calibration gas 1 2 Maintenance by service technicians This section records all changes and measures carried out by service techn...

Страница 7: ...s with the opening of the calibration lock and the implementation of a basic calibration select tab Detail 01 GC 9300 Mode Basic Calibration Following basic calibration the new response factors tab De...

Страница 8: ...Table section 1 PGC9300 Maintenance Logbook EN07 2018 June 26th 4 Table Section 1 Cylinder pressures To be performed at Every inspection Every maintenance Entries made by Operator RMG Service...

Страница 9: ...ction 1 cylinder pressures PGC9300 Maintenance Logbook EN07 2018 June 26th 5 Cylinder changed Device faulty Carriergas I Int calibration gas Carrier gas II Date Initials Pressure bar Pressure bar Pres...

Страница 10: ...ction 1 cylinder pressures PGC9300 Maintenance Logbook EN07 2018 June 26th 6 Cylinder changed Device faulty Carriergas I Int calibration gas Carrier gas II Date Initials Pressure bar Pressure bar Pres...

Страница 11: ...ction 1 cylinder pressures PGC9300 Maintenance Logbook EN07 2018 June 26th 7 Cylinder changed Device faulty Carriergas I Int calibration gas Carrier gas II Date Initials Pressure bar Pressure bar Pres...

Страница 12: ...nance Logbook EN07 2018 June 26th 8 Table section 2 Maintenance measures general To be performed at Annual routine maintenance Software update Metrological check Re calibration Repair Entries made by...

Страница 13: ...k EN07 2018 June 26th 9 Maintenance measure Performed by Performed on Measures performed Routine maintenance Table 3 Software update Metrological check Table 4 Re calibration Table 4 Defect See below...

Страница 14: ...EN07 2018 June 26th 10 Maintenance measure Performed by Performed on Measures performed Routine maintenance Table 3 Software update Metrological check Table 4 Re calibration Table 4 Defect See below...

Страница 15: ...EN07 2018 June 26th 11 Maintenance measure Performed by Performed on Measures performed Routine maintenance Table 3 Software update Metrological check Table 4 Re calibration Table 4 Defect See below...

Страница 16: ...EN07 2018 June 26th 12 Maintenance measure Performed by Performed on Measures performed Routine maintenance Table 3 Software update Metrological check Table 4 Re calibration Table 4 Defect See below...

Страница 17: ...EN07 2018 June 26th 13 Maintenance measure Performed by Performed on Measures performed Routine maintenance Table 3 Software update Metrological check Table 4 Re calibration Table 4 Defect See below...

Страница 18: ...EN07 2018 June 26th 14 Maintenance measure Performed by Performed on Measures performed Routine maintenance Table 3 Software update Metrological check Table 4 Re calibration Table 4 Defect See below...

Страница 19: ...EN07 2018 June 26th 15 Maintenance measure Performed by Performed on Measures performed Routine maintenance Table 3 Software update Metrological check Table 4 Re calibration Table 4 Defect See below...

Страница 20: ...EN07 2018 June 26th 16 Maintenance measure Performed by Performed on Measures performed Routine maintenance Table 3 Software update Metrological check Table 4 Re calibration Table 4 Defect See below...

Страница 21: ...EN07 2018 June 26th 17 Maintenance measure Performed by Performed on Measures performed Routine maintenance Table 3 Software update Metrological check Table 4 Re calibration Table 4 Defect See below...

Страница 22: ...EN07 2018 June 26th 18 Maintenance measure Performed by Performed on Measures performed Routine maintenance Table 3 Software update Metrological check Table 4 Re calibration Table 4 Defect See below...

Страница 23: ...able section 3 PGC9300 Maintenance Logbook EN07 2018 June 26th 19 Table section 3 Annual routine maintenance To be performed at Annual routine maintenance Entries made by Trained personnel RMG Service...

Страница 24: ...ck of the second pressure regulation level porter regulator if any Leak check Checklist for measuring element and GC 9300 Verification of the prefilters on the gas connector unit change of filter as n...

Страница 25: ...ck of the second pressure regulation level porter regulator if any Leak check Checklist for measuring element and GC 9300 Verification of the prefilters on the gas connector unit change of filter as n...

Страница 26: ...ck of the second pressure regulation level porter regulator if any Leak check Checklist for measuring element and GC 9300 Verification of the prefilters on the gas connector unit change of filter as n...

Страница 27: ...ck of the second pressure regulation level porter regulator if any Leak check Checklist for measuring element and GC 9300 Verification of the prefilters on the gas connector unit change of filter as n...

Страница 28: ...ck of the second pressure regulation level porter regulator if any Leak check Checklist for measuring element and GC 9300 Verification of the prefilters on the gas connector unit change of filter as n...

Страница 29: ...ck of the second pressure regulation level porter regulator if any Leak check Checklist for measuring element and GC 9300 Verification of the prefilters on the gas connector unit change of filter as n...

Страница 30: ...ck of the second pressure regulation level porter regulator if any Leak check Checklist for measuring element and GC 9300 Verification of the prefilters on the gas connector unit change of filter as n...

Страница 31: ...ck of the second pressure regulation level porter regulator if any Leak check Checklist for measuring element and GC 9300 Verification of the prefilters on the gas connector unit change of filter as n...

Страница 32: ...ck of the second pressure regulation level porter regulator if any Leak check Checklist for measuring element and GC 9300 Verification of the prefilters on the gas connector unit change of filter as n...

Страница 33: ...ck of the second pressure regulation level porter regulator if any Leak check Checklist for measuring element and GC 9300 Verification of the prefilters on the gas connector unit change of filter as n...

Страница 34: ...Table section 4 PGC9300 Maintenance Logbook EN07 2018 June 26th 30 Table section 4 Test gas analyses To be performed at Metrological check Re calibration Entries made by Trained personnel RMG Service...

Страница 35: ...nalysis of the test gases 3 Verification of the chromatograms data transmission Current outputs only need to be checked if they are used for official transmission of fiscal metering readings 4 Verific...

Страница 36: ...elevant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Ne...

Страница 37: ...ant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Neopen...

Страница 38: ...elevant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Ne...

Страница 39: ...gas 3 Verification of the chromatogram of test gas 4 Verification of the chromatogram of test gas 5 Verification of the chromatogram of test gas 6 Checking the range of the current outputs Verificati...

Страница 40: ...2 Analysis of the test gases 3 Verification of the chromatograms data transmission Current outputs only need to be checked if they are used for official transmission of fiscal metering readings 4 Veri...

Страница 41: ...ant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Neopen...

Страница 42: ...elevant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Ne...

Страница 43: ...ant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Neopen...

Страница 44: ...est gas 3 Verification of the chromatogram of test gas 4 Verification of the chromatogram of test gas 5 Verification of the chromatogram of test gas 6 Checking the range of the current outputs Verific...

Страница 45: ...nalysis of the test gases 3 Verification of the chromatograms data transmission Current outputs only need to be checked if they are used for official transmission of fiscal metering readings 4 Verific...

Страница 46: ...elevant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Ne...

Страница 47: ...ant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Neopen...

Страница 48: ...elevant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Ne...

Страница 49: ...gas 3 Verification of the chromatogram of test gas 4 Verification of the chromatogram of test gas 5 Verification of the chromatogram of test gas 6 Checking the range of the current outputs Verificati...

Страница 50: ...2 Analysis of the test gases 3 Verification of the chromatograms data transmission Current outputs only need to be checked if they are used for official transmission of fiscal metering readings 4 Veri...

Страница 51: ...ant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Neopen...

Страница 52: ...elevant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Ne...

Страница 53: ...ant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Neopen...

Страница 54: ...est gas 3 Verification of the chromatogram of test gas 4 Verification of the chromatogram of test gas 5 Verification of the chromatogram of test gas 6 Checking the range of the current outputs Verific...

Страница 55: ...nalysis of the test gases 3 Verification of the chromatograms data transmission Current outputs only need to be checked if they are used for official transmission of fiscal metering readings 4 Verific...

Страница 56: ...elevant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Ne...

Страница 57: ...ant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Neopen...

Страница 58: ...elevant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Ne...

Страница 59: ...gas 3 Verification of the chromatogram of test gas 4 Verification of the chromatogram of test gas 5 Verification of the chromatogram of test gas 6 Checking the range of the current outputs Verificati...

Страница 60: ...2 Analysis of the test gases 3 Verification of the chromatograms data transmission Current outputs only need to be checked if they are used for official transmission of fiscal metering readings 4 Veri...

Страница 61: ...ant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Neopen...

Страница 62: ...elevant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Ne...

Страница 63: ...ant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Neopen...

Страница 64: ...est gas 3 Verification of the chromatogram of test gas 4 Verification of the chromatogram of test gas 5 Verification of the chromatogram of test gas 6 Checking the range of the current outputs Verific...

Страница 65: ...nalysis of the test gases 3 Verification of the chromatograms data transmission Current outputs only need to be checked if they are used for official transmission of fiscal metering readings 4 Verific...

Страница 66: ...elevant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Ne...

Страница 67: ...ant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Neopen...

Страница 68: ...elevant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Ne...

Страница 69: ...gas 3 Verification of the chromatogram of test gas 4 Verification of the chromatogram of test gas 5 Verification of the chromatogram of test gas 6 Checking the range of the current outputs Verificati...

Страница 70: ...2 Analysis of the test gases 3 Verification of the chromatograms data transmission Current outputs only need to be checked if they are used for official transmission of fiscal metering readings 4 Veri...

Страница 71: ...ant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Neopen...

Страница 72: ...elevant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Ne...

Страница 73: ...ant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Neopen...

Страница 74: ...est gas 3 Verification of the chromatogram of test gas 4 Verification of the chromatogram of test gas 5 Verification of the chromatogram of test gas 6 Checking the range of the current outputs Verific...

Страница 75: ...nalysis of the test gases 3 Verification of the chromatograms data transmission Current outputs only need to be checked if they are used for official transmission of fiscal metering readings 4 Verific...

Страница 76: ...elevant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Ne...

Страница 77: ...ant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Neopen...

Страница 78: ...elevant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Ne...

Страница 79: ...gas 3 Verification of the chromatogram of test gas 4 Verification of the chromatogram of test gas 5 Verification of the chromatogram of test gas 6 Checking the range of the current outputs Verificati...

Страница 80: ...2 Analysis of the test gases 3 Verification of the chromatograms data transmission Current outputs only need to be checked if they are used for official transmission of fiscal metering readings 4 Veri...

Страница 81: ...ant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Neopen...

Страница 82: ...elevant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Ne...

Страница 83: ...ant for calibration Test gas 1 Test gas 2 Type Error Type Error set actual absolute set actual absolute Calorific value Standard density Carbon dioxide Methane Ethane Propane Isobutane N butane Neopen...

Страница 84: ...est gas 3 Verification of the chromatogram of test gas 4 Verification of the chromatogram of test gas 5 Verification of the chromatogram of test gas 6 Checking the range of the current outputs Verific...

Страница 85: ...For further information please visit our website www rmg com or contact your local sales support office to learn more about the RMG products RMG Messtechnik GmbH Otto Hahn Stra e 5 35510 Butzbach Ger...

Отзывы: