Maintenance and service
In order to enable a long and trouble-free operation of the analyzer, maintenance on the device
must be carried out at regular intervals. Make sure the device is turned off before performing any
maintenance. During this time no analyses are carried out. As a general rule, wear protective
goggles and gloves during maintenance to avoid contact with reagent, cleaning fluid or other
liquids.
The following maintenance intervals must be complied with:
Interval
Maintenance and service
every 6 months
Cleaning of measurement chamber
At high ambient- and water temperatures or
water with high organic load, the cleaning
intervals may need to be shortened.
every 30,000 analyses or after 24 months
maintenance as after 6 months and
installing maintenance kit
For cleaning, we recommend the "SYCON Clean Cleaning Set" See page 70.