70
Annex
Maintenance sets
The SYCON 2702 is largely maintenance-free. A maintenance set is available for the analyzer. It is
recommended to change the hose pump cassette, hoses and O-rings after 30,000 analyses or 24
months. At the end of the maintenance interval, the instrument displays a maintenance note on the
display. The maintenance counter must be reset or confirm after maintenance has been carried out.
In addition, it is recommended to clean the measuring chamber regularly, but at least every 6
months. The SYCON Clean cleaning set is offered for this purpose. It contains all the aids required
for cleaning as well as the cleaning fluid FIT 3000.
Information on performing maintenance can be found in the chapter "Maintenance and Service"
starting on page 55.
Article name
Article no.
Maintenance set for SYCON 2702
Includes the following articles:
•
1 x 33-090 008 bottle connector
•
1 x 33-090 011 suction lance
•
1 x 33-090 217 O-ring 16x2
•
4 x 33-090 210 O-ring 9x1.5
•
1 x 33-090 218 O-ring blue
•
1 x 33-030 614 peristaltic pump
cartridge
•
1 x 33-090002 Agitator blade
33-090 034
SYCON Clean cleaning set
Cleaning set for measuring chambers,
contains the following items:
•
10 pairs of gloves XL
•
1 x funnel made of PP
•
1 x cleaner FIT 3000 (1000 ml)
•
1 x pipette brush
•
1 x reagent glass brush
•
1 x container with lid
30-010 900
Measuring chamber cleaner FIT 3000 (1000 ml)
32-089 100