Auto. Time interval
Set the automatic interval time, in which the analyses are to
be carried out.
You will see an input keyboard on the right side.
Use the arrow keys ([
], [
], [
] and [
]) to move the black
cursor to the desired digit and press the [OK] key.
The selected digit appears on the left side in a frame.
Repeat the entry until the desired analysis interval time is
within the range.
Now move the cursor to the OK field in the keyboard and
press the [OK] key.
You will see the following information on the left:
Current:
currently programmed value
Minimum:
smallest settable value
Maximum:
largest settable value
Should the value at "Current" corresponds to your request,
then you do not need to re-enter the number and you can
immediately move the cursor to the OK field in the keyboard
and confirm with the [OK] key.
Control measurement
How often should a limit exceedance be verified before
notification?
Press [OK] key.
Control measurement
Use [
] and [
] to set the number of control measurements
that should be performed before any exceedance of limit
value should be reported. These are carried out after a limit
value exceedance in a 3-minute interval to avoid false alarms
due to the counter-ion effect of the water softening system.
Make a selection and confirm with the [OK] key.
Configuration