Safety reliably produced.
© RLS Wacon analytics GmbH ∙ +49 (0)5121 28126 0 ∙
Subject to
change without notice
59
Maintenance and service
Diagnostic functions LIMESPlus
The diagnostic programme is used to check the unit functions. This is available in the LIMESPlus
version.
This test may only be called up by qualified persons. Please observe the following safety
instructions!
Make a note of the positions of the programme switches before the diagnosis and
switch all programme switches back to the original position after the diagnostic
programme has been carried out.
Use the service booklet to note down your settings.
The test must be carried out with the front cover open, and the rubber cap removed. All buttons
and switches are operated.
When running the diagnostic programme, please note the following:
•
Depending on the wiring, the cables of relays 1 and 2 exiting the LIMES may be
connected to mains voltage.
•
Touching it can lead to serious injuries.
•
There is danger to life.
•
The unit can be damaged by improper handling.
Activation of relay 1 and relay 2 can lead to operating faults.
Note that the supply line is under pressure.
Avoid splashing water that could get into the electronics and destroy them.
Switch SW4 to the ON position. The diagnostic functions described below can then be called up
one after the other. The unit is in diagnostic step 1 at the beginning.
Programme switch
SW4
Function
ON
Diagnostic programme
Each time the "Start" key is pressed, the system switches to the next diagnostic step.
Voltage
Pressure