65-2437RKSS-05 CO Detector • 3

65-2438 CO Detector

The 65-2438 CO detector consists of the detector housing body, detector housing cap, cap 
gasket, the plug-in sensor, and the charcoal filter with rubber boot. 

Figure 2: CO Detector Component Location

Detector Housing, Housing Cap, & Cap Gasket

The detector housing body protects the sensing components within the housing. Use the 
3/4” NPT mounting threads at the top of the housing to screw the CO detector into the 
bottom conduit hubs of the junction box. Use the removable cap near the bottom of the 
housing to access the sensor for maintenance or replacement. The cap protects the sensor 
from damage and includes a flame arrestor which contains any sparks which may occur 
within the detector housing. A cap gasket seals the interface between the housing and cap. 
A flame arrestor guard is permanently bonded to the cap.

Two wires extend from the top of the detector housing body. Use these wires to connect 
the CO detector to the junction box terminal strip. The housing includes a four-socket 
pattern. This socket pattern accepts the sensor’s four pins to secure the sensor within the 
detector housing. A pre-amplifier, located between the sockets and two interconnect 
wires, conditions the sensor’s signal before the signal reaches the controller.

Plug-In Sensor

The sensor is secured within the sensor housing by four pins. Through a series of chemical 
and electrical reactions, the sensor produces an electrical output that corresponds to the 
detection range of the detector.

Charcoal Filter

The disc-shaped charcoal filter is secured to the face of the CO sensor with a rubber boot. 
The charcoal filter prevents interference gases (hydrogen sulfide [H

2

S] and certain 

hydrocarbons) from producing false CO readings.

Detector Housing Cap

Rubber Boot With 
Charcoal Filter

Plug-In CO Sensor

Cap Gasket

Detector Housing Body

Flame Arrestor 
Guard

Содержание 65-2437RKSS-05

Страница 1: ...RKI Instruments Inc www rkiinstruments com 65 2437RKSS 05 Carbon Monoxide Detector Operator s Manual Part Number 71 0439 Revision P1 Released 9 14 17...

Страница 2: ...calibration and maintenance of the detector is essential for proper operation and correct readings Please calibrate and maintain this detector regularly Frequency of calibration depends upon the type...

Страница 3: ...t costs local repair costs transportation costs or contingent expenses incurred without our prior approval THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND REPRESENTATIONS EXPRES...

Страница 4: ...he CO Detector 4 Wiring the CO Detector to a Controller 5 Startup 7 Introducing Incoming Power 7 Setting the Zero Signal 7 Maintenance 8 Preventive Maintenance 8 Troubleshooting 9 Replacing Components...

Страница 5: ...e instructions and warnings in this manual to assure proper and safe operation of the 65 2437RKSS 05 and to minimize the risk of personal injury Be sure to maintain and periodically calibrate the 65 2...

Страница 6: ...on describes the components of the 65 2437RKSS 05 CO detector The 65 2437RKSS 05 consists of the 65 2438 CO detector and junction box Figure 1 CO Detector Component Location Stainless Steel J Box Lock...

Страница 7: ...ntly bonded to the cap Two wires extend from the top of the detector housing body Use these wires to connect the CO detector to the junction box terminal strip The housing includes a four socket patte...

Страница 8: ...the junction box A locking set screw on the junction box cover allows you to secure the junction box cover and prevent it from being removed Installation This section describes procedures to mount th...

Страница 9: ...source is OFF before you make wiring connections 1 Turn off the controller 2 Turn off or unplug power to the controller 3 Remove the junction box cover WARNING To maintain the explosion proof classif...

Страница 10: ...junction box cover to the junction box 7 Route the cable or wires leading from the CO detector through one of the conduit hubs at the controller housing Use appropriate conduit fittings and constructi...

Страница 11: ...the Zero Fresh Air Reading WARNING If the detector installation is explosion proof do not remove the detector housing cap or junction box cover while the circuits are energized unless the area is dete...

Страница 12: ...ble external alarms 2 Verify that the controller display reading for the channel you are testing is 0 If the display reading is not zero set the zero reading of the detector as described in the Start...

Страница 13: ...lt or Unable to Calibrate Detector responds slowly or does not respond during response test Unable to accurately set the zero or response reading during calibration Detector requires frequent calibrat...

Страница 14: ...in the detector housing 8 Make sure the cap gasket is in place and screw the detector housing cap back onto the detector housing body 9 Turn on or plug in power to the controller 10 Turn on the contro...

Страница 15: ...ect 8 Reinstall the junction box cover 9 Turn on or plug in power to the controller 10 Turn on the controller and place into normal operation CAUTION Allow the replacement detector to warm up for 5 mi...

Страница 16: ...he fixed flow regulator to the calibration cup 4 Put the controller into its calibration program Setting the Zero Reading NOTE If you can verify that the CO detector is in a fresh air environment you...

Страница 17: ...regulator from the cylinder 8 Remove the calibration cup from the detector NOTE For convenience leave regulator and calibration cup connected by the sample tubing 9 Allow about 45 seconds for the gas...

Страница 18: ...9 65 2437RKSS 05 CO Detector Operator s Manual this document 81 0064RK Calibration cylinder 50 PPM CO in air 17 liter 81 0064RK 01 Calibration cylinder 50 PPM CO in air 34 liter steel 81 0064RK 03 Cal...

Отзывы: