background image

 

65-2331-04 H

2

S Transmitter • 5

Zero Pot

The zero pot is located in the upper left corner of the amplifier (see Figure 3). Use a small 
flat blade screwdriver to turn the zero pot’s adjustment screw and adjust the amplifier’s 
zero (H

2

S free) output during the start-up and calibration procedures. Turn the 

adjustment screw clockwise to increase the zero output and counterclockwise to decrease 
the zero output.

Span Pot

The span pot is located to the right of the zero pot (see Figure 3). Use a small flat blade 
screwdriver to turn the span pot’s adjustment screw and adjust the amplifier’s response 
output during the calibration procedure. Turn the adjustment screw clockwise to increase 
the output and counterclockwise to decrease the output.

CAUTION:

The third potentiometer is factory-set. Do not adjust it.

Test Points

The test points (red and black) are on the left side of the amplifier (see Figure 3). The test 
points produce a 100mV to 500 mV output that corresponds to the transmitter’s 4 to 20 
mA output. Use the test points and a voltmeter to measure the transmitter’s output 
during the start-up and calibration procedures. The black test point in the lower left 
corner is the negative (-) test point and the red test point below the zero pot is the positive 
(+) test point.

Amplifier Junction Box

The amplifier junction box protects the amplifier and wiring connections made to the 
amplifier. Use the top 3/4” conduit hub to connect wiring from the amplifier to the 
controller. Use the bottom 3/4” conduit hub to wire the remotely installed H

2

S detector. 

Use the cover on the front of the junction box to access the interior of the junction box. The 
amplifier is factory installed in the junction box. Three spacers installed on the back of the 
junction box controls the distance of the junction box from a mounting surface.

Detector Junction Box

The detector junction box is intended to be installed remotely from the amplifier junction 
box. The H

2

S detector is factory installed in the bottom 3/4” conduit hub. A terminal strip 

is provided for wiring connections and the detector is factory wired to one side of this 
terminal strip. The other side of the terminal strip is for user-supplied wiring to the 
detector terminal strip located in the amplifier junction box. This wiring should go 
through the top 3/4” conduit hub. Three spacers installed on the back of the junction box 
controls the distance of the junction box from a mounting surface.

Содержание 65-2331-04

Страница 1: ...RKI Instruments Inc www rkiinstruments com 65 2331 04 Hydrogen Sulfide Transmitter Operator s Manual Part Number 71 0440 Revision P1 Released 9 15 17...

Страница 2: ...alibration and maintenance of the detector is essential for proper operation and correct readings Please calibrate and maintain this detector regularly Frequency of calibration depends upon the type o...

Страница 3: ...t costs local repair costs transportation costs or contingent expenses incurred without our prior approval THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND REPRESENTATIONS EXPRES...

Страница 4: ...the Detector to the Amplifier 8 Wiring the Amplifier to a Controller 9 Startup 11 Introducing Incoming Power 11 Setting the Zero Signal 12 Maintenance 13 Preventive Maintenance 13 Troubleshooting 15 R...

Страница 5: ...ual to assure proper and safe operation of the 65 2331 04 and to minimize the risk of personal injury Be sure to maintain and periodically calibrate the 65 2331 04 as described in this manual Table 1...

Страница 6: ...ox The two junction box configuration is intended for situations where the detector needs to be installed at an inaccessible location The detector junction box can be installed at the inaccessible loc...

Страница 7: ...e sensor within the detector housing A pre amplifier located between the sockets and two interconnect wires conditions the sensor s signal before the signal reaches the amplifier Housing Cap Cap Gaske...

Страница 8: ...the sensor to a 4 to 20 mA signal that corresponds to the detection range and transmits the signal to a gas monitoring controller A foam gasket that orients the amplifier and keeps it from rotating i...

Страница 9: ...ures The black test point in the lower left corner is the negative test point and the red test point below the zero pot is the positive test point Amplifier Junction Box The amplifier junction box pro...

Страница 10: ...isturbed Make sure there is sufficient room to perform start up maintenance and calibration procedures The site should be easily accessible For the detector junction box select a site where the target...

Страница 11: ...see Figure 5 4 Install 3 16 inch I D flexible polyurethane tubing to the fitting at the bottom of the calibration adapter splash guard and route it to an accessible area that is close to the amplifier...

Страница 12: ...drain wire disconnected and insulated at the detector junction box You will connect the opposite end of the cable s drain wire to the amplifier junction box s chassis earth ground 6 Secure the detecto...

Страница 13: ...ient grounding location Install a lug on the shield drain wire or wrap the shield drain wire around the mounting screw Do not connect the shield drain wire at the detector junction box Wiring the Ampl...

Страница 14: ...used leave the cable shield s drain wire disconnected and insulated at the amplifier junction box You will connect the opposite end of the cable s drain wire to the controller s chassis earth ground...

Страница 15: ...s procedures to start up the H2S transmitter and place the transmitter into normal operation Introducing Incoming Power 1 Complete the installation procedures described earlier in this manual 2 Verify...

Страница 16: ...elow describes applying zero emission air usually called zero air using a calibration kit that includes calibration gas sample tubing and a fixed flow regulator with an on off knob RKI Instruments Inc...

Страница 17: ...the H2S transmitter may cause alarms Be sure to put the controller into its calibration program or disable external alarms before performing this test Preparing for the response test 1 Place the contr...

Страница 18: ...e reading calibrate the affected transmitter s as described in the Calibration on page 19 4 Turn the regulator s on off knob clockwise to close the regulator 5 Unscrew the regulator from the calibrati...

Страница 19: ...slowly or does not respond to response test Unable to accurately set the zero or response reading during calibration Transmitter requires frequent calibration Note Under normal circumstances the trans...

Страница 20: ...r housing 8 Make sure the cap gasket is in place and screw the detector housing cap back onto the detector housing body 9 Turn on or plug in power to the controller 10 Turn on the controller and place...

Страница 21: ...controller terminal strip from their sockets You may leave the wires connected to the terminal strips 5 Unscrew and remove the screw with the flat and lock washers that secures the amplifier to the j...

Страница 22: ...at is attached to the calibration adapter splash guard To determine the response time based on the length of tubing connected 1 Place the controller into its calibration mode or disable external alarm...

Страница 23: ...on the amplifier Plug the positive lead into the red test point plug the negative lead into the black test point 4 Use the following formula to determine the correct test points output for the calibra...

Страница 24: ...termined earlier If necessary use the span pot on the amplifier to adjust the reading to match the correct response reading 5 Turn the regulator knob clockwise to close the regulator 6 Unscrew the reg...

Страница 25: ...udes detector two junction boxes and amplifier CSA classified 65 2428 H2S replacement detector assembly includes sensor CSA classified 71 0440 65 2331 04 H2S Transmitter Operator s Manual this documen...

Отзывы: