18 • 65-2331-04 H

2

S Transmitter

13. Allow the transmitter to warm up for 5 minutes.

14. Calibrate the H

2

S transmitter as described in the “Calibration” on page 19.

Calibration Frequency

Although there is no particular calibration frequency that is correct for all applications, a 
calibration frequency of every 3 months is adequate for most H

2

S transmitter applications. 

Unless experience in a particular application dictates otherwise, RKI Instruments, Inc. 
recommends a calibration frequency of every 3 months for the H

2

S transmitter.

If an application is not very demanding, for example detection in a clean, temperature 
controlled environment, and calibration adjustments are minimal at calibration, then a 
calibration frequency of every 6 months is adequate.

If the application is very demanding, for example if the environment is not well 
controlled, then more frequent calibration than every 3 months may be necessary.

Determining Response Time

Since the detector junction box is installed remotely from the amplifier, the response time 
of the detector will depend on the length of flexible tubing that is attached to the 
calibration adapter/splash guard. To determine the response time based on the length of 
tubing connected:

1.

Place the controller into its calibration mode or disable external alarms.

2.

Note the current gas reading displayed at the controller.

3.

Screw the regulator into the calibration cylinder. Do not use a zero air cylinder for this 
operation.

4.

Use the flexible tubing coming from the calibration adapter/splash guard to connect 
the regulator to the calibration cup.

5.

Turn the regulator’s on/off knob counterclockwise to open the regulator. Gas will 
begin to flow. Take note of the time or start a stopwatch.

6.

When you first start to notice an increase in the gas reading, note how much time 
passed between starting the gas flow and noticing a reading response.

7.

Turn the regulator’s on/off knob clockwise to close the regulator.

8.

Unscrew the regulator from the calibration cylinder.

9.

Disconnect the calibration adapter/splash guard’s flexible tubing from the regulator.

NOTE:

Do not disconnect the flexible tubing from the calibration adapter/splash guard.

10. Add 1 minute to the length of time determined in Step 6. This is the response time for 

the detector. When performing a response test, zero adjustment, or span adjustment, 
gas must be applied for this length of time.

Содержание 65-2331-04

Страница 1: ...RKI Instruments Inc www rkiinstruments com 65 2331 04 Hydrogen Sulfide Transmitter Operator s Manual Part Number 71 0440 Revision P1 Released 9 15 17...

Страница 2: ...alibration and maintenance of the detector is essential for proper operation and correct readings Please calibrate and maintain this detector regularly Frequency of calibration depends upon the type o...

Страница 3: ...t costs local repair costs transportation costs or contingent expenses incurred without our prior approval THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND REPRESENTATIONS EXPRES...

Страница 4: ...the Detector to the Amplifier 8 Wiring the Amplifier to a Controller 9 Startup 11 Introducing Incoming Power 11 Setting the Zero Signal 12 Maintenance 13 Preventive Maintenance 13 Troubleshooting 15 R...

Страница 5: ...ual to assure proper and safe operation of the 65 2331 04 and to minimize the risk of personal injury Be sure to maintain and periodically calibrate the 65 2331 04 as described in this manual Table 1...

Страница 6: ...ox The two junction box configuration is intended for situations where the detector needs to be installed at an inaccessible location The detector junction box can be installed at the inaccessible loc...

Страница 7: ...e sensor within the detector housing A pre amplifier located between the sockets and two interconnect wires conditions the sensor s signal before the signal reaches the amplifier Housing Cap Cap Gaske...

Страница 8: ...the sensor to a 4 to 20 mA signal that corresponds to the detection range and transmits the signal to a gas monitoring controller A foam gasket that orients the amplifier and keeps it from rotating i...

Страница 9: ...ures The black test point in the lower left corner is the negative test point and the red test point below the zero pot is the positive test point Amplifier Junction Box The amplifier junction box pro...

Страница 10: ...isturbed Make sure there is sufficient room to perform start up maintenance and calibration procedures The site should be easily accessible For the detector junction box select a site where the target...

Страница 11: ...see Figure 5 4 Install 3 16 inch I D flexible polyurethane tubing to the fitting at the bottom of the calibration adapter splash guard and route it to an accessible area that is close to the amplifier...

Страница 12: ...drain wire disconnected and insulated at the detector junction box You will connect the opposite end of the cable s drain wire to the amplifier junction box s chassis earth ground 6 Secure the detecto...

Страница 13: ...ient grounding location Install a lug on the shield drain wire or wrap the shield drain wire around the mounting screw Do not connect the shield drain wire at the detector junction box Wiring the Ampl...

Страница 14: ...used leave the cable shield s drain wire disconnected and insulated at the amplifier junction box You will connect the opposite end of the cable s drain wire to the controller s chassis earth ground...

Страница 15: ...s procedures to start up the H2S transmitter and place the transmitter into normal operation Introducing Incoming Power 1 Complete the installation procedures described earlier in this manual 2 Verify...

Страница 16: ...elow describes applying zero emission air usually called zero air using a calibration kit that includes calibration gas sample tubing and a fixed flow regulator with an on off knob RKI Instruments Inc...

Страница 17: ...the H2S transmitter may cause alarms Be sure to put the controller into its calibration program or disable external alarms before performing this test Preparing for the response test 1 Place the contr...

Страница 18: ...e reading calibrate the affected transmitter s as described in the Calibration on page 19 4 Turn the regulator s on off knob clockwise to close the regulator 5 Unscrew the regulator from the calibrati...

Страница 19: ...slowly or does not respond to response test Unable to accurately set the zero or response reading during calibration Transmitter requires frequent calibration Note Under normal circumstances the trans...

Страница 20: ...r housing 8 Make sure the cap gasket is in place and screw the detector housing cap back onto the detector housing body 9 Turn on or plug in power to the controller 10 Turn on the controller and place...

Страница 21: ...controller terminal strip from their sockets You may leave the wires connected to the terminal strips 5 Unscrew and remove the screw with the flat and lock washers that secures the amplifier to the j...

Страница 22: ...at is attached to the calibration adapter splash guard To determine the response time based on the length of tubing connected 1 Place the controller into its calibration mode or disable external alarm...

Страница 23: ...on the amplifier Plug the positive lead into the red test point plug the negative lead into the black test point 4 Use the following formula to determine the correct test points output for the calibra...

Страница 24: ...termined earlier If necessary use the span pot on the amplifier to adjust the reading to match the correct response reading 5 Turn the regulator knob clockwise to close the regulator 6 Unscrew the reg...

Страница 25: ...udes detector two junction boxes and amplifier CSA classified 65 2428 H2S replacement detector assembly includes sensor CSA classified 71 0440 65 2331 04 H2S Transmitter Operator s Manual this documen...

Отзывы: