37
Mode
Item
LCD display
Details
Maintenance
mode
(Regular
maintenance)
Gas introduction
display
2-0 GAS TEST
2-00: Gas Test
2-01: Alarm Test
2-02: Fault Alarm Test
2-03: Display test
2-04: Not used (----)
Zero adjustment
2-1 ZERO
Perform the zero adjustment.
Span adjustment
2-2 SPAN
Perform the span adjustment.
Zero/span
initialization
2-3 SDEF
Not used.
Environmental
setting
2-4 SET
2-40: Not used (----)
2-41: Not used (----)
2-42: Alarm Value Setting (AL-P)
2-43: Alarm Delay Time Setting (AL-D)
2-44: Alarm Pattern Setting (AL-T)
2-45: Zero Suppression Type Setting (SP-T)
2-46: Zero Suppression Value Setting (SAPP)
2-47: Not used (----)
2-48: External Output Adjustment (MA20)
2-49: Not used (----)
2-4A: Date/Time Setting (DATE)
2-4b: Not used (----)
2-4C: Not used (----)
2-4d: Buzzer ON/OFF Setting (BZZR)
2-4E: Gas Alarm Contact ON/OFF (AL-R)
2-4F: External Output Signal ON/OFF
2-4H: LCD Backlight ON/OFF (LCD)
Display
2-5 DISP
2-50
:
Not used (----)
2-51
:
Not used (----)
2-52
:
Calibration Curve Number Display (GSEL)
2-53
:
Fault Detail Display (FALT)
2-54
:
Not used (----)
Factory mode
switching
2-6 F MODE
Not used.
User mode
switching
2-7 U MODE
Return to the user mode.