54
<Replacing the Sensor (Remote Sensor Type)>
1
Remove the sensor cover of the
sensor unit.
Push the tabs on the both sides with
your fingers to release the sensor
cover latch.
Remove the sensor cover to check
the sensor.
2
Remove the old sensor.
3
Install the new sensor.
Be sure the sensor’s pins are lined
up with the board’s sockets.
4
Put the sensor cover back on.
7-9. Replacing a Sensor Cable
NOTE: The wire colors in the extender cable may vary depending on the type of cable used. The
Type A cable has red/green/white/black wires and the Type B cable has
white/yellow/white/blue wires. The pictures in this section show the Type A cable.
2
Remove the sensor cover of the old
cable’s remote sensor housing.
Push the tabs on both sides of the sensor
cover with your fingers to release the sensor
cover latch.
3
Remove the cable plug from the
cable socket on the sensor board.
Press the tab on the plug to release it and
pull it away from the socket.
1
Turn off or disconnect power to the
EC-600.