04 Series for CO(-H2)
Appendix B: Gas Select Mode • 89
7. The
SAVE-AP
item displays.
8. Use AIR to move through the Gas Select Mode items.
Tips for Using Gas Select Mode
•
To scroll from one item to the next, press and release AIR. To reverse the scrolling
direction:
a. Press and hold AIR.
b. Immediately press POWER MODE and then release both buttons.
c. The scrolling direction returns to the original direction when you exit and reenter a
menu.
•
To skip an item when a question is asked, press and release AIR.
•
To enter an item and to save any changes, press and release POWER MODE.
•
To change a flashing parameter, press and release AIR. To reverse the direction of change
(ie. from increasing to decreasing or vice versa):
a. Press and hold AIR.
b. Immediately press POWER MODE and then release both buttons.
•
To exit an entered item without saving a change, press and hold AIR and POWER MODE
for a few seconds.
Saving the Alarm Points (SAVE-AP)
Performing a
SAVE-AP
operation saves the current alarm setpoints.
Performing a
DEF.ALMP
operation in the
ALARM-P
User Mode menu sets the instrument’s
alarm points to those saved during the
SAVE-AP
operation (if performed).
Performing a
SAVE-AP
has no effect on an
M.DEF
in
Maintenance Mode. An
M.DEF
operation returns all instrument settings to the RKI default regardless of if a
SAVE-AP
operation was performed.
1. While in Gas Select Mode, press AIR to scroll to
SAVE-AP
.
SAVE - AP
SAVE - AP