04 Series for CO(-H2)
Chapter 5: User Mode and Calibration • 57
1. After entering the
BUMP.SET
menu, press AIR to scroll to
BP.EXPD
.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the
BP.EXPD
item in
the
BUMP.SET
menu.
5. See “Exiting the BUMP.SET Menu” on page 57 to return to User Mode.
Exiting the BUMP.SET Menu
1. While in the
BUMP.SET
menu, press AIR to scroll to
ESCAPE
.
2. Press and release POWER MODE to return to the
BUMP.SET
item in User Mode.
3. See “Entering Measuring Mode (START)” on page 64 to enter Measuring Mode.
Alarm Settings (ALARM--P)
The
ALARM--P
menu has 3 items:
ALARM--P
,
DEF.ALMP
, and
ESCAPE
.
1. While in User Mode, press AIR to scroll to
ALARM--P
.
2. Press and release POWER MODE. The first screen displays.
Setting the Alarm Points
1. After entering the
ALARM-P
menu, press AIR to scroll to the gas name.
2. Press and release POWER MODE.
BP .EXPD
ALARM-P
CO
ALARM-P
ppm
AIR
AIR
AIR
CO
ALARM-P
ppm
DEF .ALMP
ESCAPE